Skip to content

SET Rotarex™S 8 F x 110 cm 80239Recalls & reported problems

Straub Medical AG · Rotational Excisional Atherectomy System. The Catheter Set includes one Rotarex™S catheter (gauge 8F; length 110 cm), one guidewire (0.018” diameter; 270cm in length), one drape and one collecting bag. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07640142811886), received between Oct 22, 2025 and Mar 3, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 5, 2025Class IOpen, Classified

    Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.

    Z-1244-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Deformation Due to Compressive Stress1
  • Leak/Splash1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Perforation of Vessels1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 3, 2026Leak/Splash; Deformation Due to Compressive Stress

    ON (B)(6) 2026, A PATIENT UNDERWENT A THROMBECTOMY AND ATHERECTOMY PROCEDURE USING THE ROTAREX DEVICE. DURING THE PROCEDURE, IT WAS REPORTED THAT THE BLOOD WAS NOTED TO BE LEAKING FROM THE METAL PORTION OF THE CATHETER LOCATED ABOVE THE HANDLE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24492919 ↗
  • InjuryOct 22, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    ON (B)(6) 2025, A PATIENT UNDERWENT THROMBECTOMY AND ATHERECTOMY PROCEDURE USING A ROTAREX DEVICE. IT WAS REPORTED THAT ONE 8F ROTAREX GOT STUCK ON THE WIRE. THERE WAS AN ILIAC PERF AND THEY LOST WIRE ACCESS. THE PATIENT WAS SENT TO SURGERY TO REPAIR THE ILIAC. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Perforation of Vessels

    MAUDE report 23356367 ↗

Other Catheter, Peripheral, Atherectomy devices

All Catheter, Peripheral, Atherectomy devices →

Sources and updates