Skip to content

SPI ELEMENT INICELL 4.23.554Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI ELEMENT INICELL (67)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 07640156470017), received between Feb 25, 2019 and May 28, 2026 — classified as 12 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate9
  • Loss of Osseointegration3

Patient problems

  • Inadequate Osseointegration10
  • No Code Available2

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 28, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, THREAD TAP USED, DISEASED MUCOUS MEMBRANE, PAIN, SWELLING, ABSCESS, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25315373 ↗
  • InjuryMar 21, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21662070 ↗
  • InjuryJun 28, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, THREAD TAP USED, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, IMMEDIATE IMPLANTATION, INADEQUATE BONE QUALITY/QUANTITY, PAIN, INFLAMMATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19630988 ↗
  • InjuryFeb 8, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, NO THREAD TAP USED, SMOKER, DISEASED MUCOUS MEMBRANE, IMMEDIATE IMPLANTATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18665940 ↗
  • InjuryNov 22, 2022Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY. EVEN WITH THE EXCELLENT INITIAL STABILITY, ONE OF TWO IMPLANTS FAILED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15845826 ↗
  • InjuryOct 14, 2022Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, XEROSTOMIA, HEAVY SMOKER, PERIODONTAL DISEASE, IMMEDIATE IMPLANTATION, INFECTION. SECOND TIME #27 FAILED, WHICH IS WHY RESTORATION DATE PRECEDES ABUTMENT DATE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15602934 ↗
  • InjuryJun 3, 2022Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PSYCHOLOGICAL DISORDER, BRUXISM, PAIN.

    Patient problems: Inadequate Osseointegration

    MAUDE report 14588065 ↗
  • InjuryOct 31, 2021Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED - IMPLANT WAS PLACED RIGHT AFTER EXPLANTATION OF AN OLDER IMPLANT - IMPLANT DIAMETER WAS TOO BIG.

    Patient problems: Inadequate Osseointegration

    MAUDE report 12725534 ↗
  • InjuryJun 22, 2021Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 12038684 ↗
  • InjuryFeb 5, 2021Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 11285738 ↗
  • InjuryAug 10, 2020Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED. 6.0 IMPLANT PLACED AND WAS TOO WIDE, CHANGED TO 5.0 BUT WAS GOING TO BE TOO LONG AND PLACED 5.0X9.5L. SAME PATIENT AS (B)(6).

    Patient problems: No Code Available

    MAUDE report 10383685 ↗
  • InjuryFeb 25, 2019Failure to Osseointegrate; Failure to Osseointegrate

    NO OSSEOINTEGRATION, THREAD TAP USED, IMPLANT WAS COMPLETELY COVERED BY BONE, IMMEDIATE IMPLANTATION, INFECTION, MOBILITY CHECK-UP AND LASER DONE THE SAME DAY.

    Patient problems: No Code Available

    MAUDE report 8365913 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page