SPI ELEMENT INICELL 4.23.554Recalls & reported problems
Thommen Medical AG · Class 2 · Implant
All models of SPI ELEMENT INICELL (67)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 07640156470017), received between Feb 25, 2019 and May 28, 2026 — classified as 12 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate9
- Loss of Osseointegration3
Patient problems
- Inadequate Osseointegration10
- No Code Available2
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
InjuryMay 28, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, THREAD TAP USED, DISEASED MUCOUS MEMBRANE, PAIN, SWELLING, ABSCESS, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 25315373 ↗InjuryMar 21, 2025Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED.
Patient problems: Inadequate Osseointegration
MAUDE report 21662070 ↗InjuryJun 28, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, THREAD TAP USED, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, IMMEDIATE IMPLANTATION, INADEQUATE BONE QUALITY/QUANTITY, PAIN, INFLAMMATION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 19630988 ↗InjuryFeb 8, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, NO THREAD TAP USED, SMOKER, DISEASED MUCOUS MEMBRANE, IMMEDIATE IMPLANTATION.
Patient problems: Inadequate Osseointegration
MAUDE report 18665940 ↗InjuryNov 22, 2022Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY. EVEN WITH THE EXCELLENT INITIAL STABILITY, ONE OF TWO IMPLANTS FAILED.
Patient problems: Inadequate Osseointegration
MAUDE report 15845826 ↗InjuryOct 14, 2022Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, XEROSTOMIA, HEAVY SMOKER, PERIODONTAL DISEASE, IMMEDIATE IMPLANTATION, INFECTION. SECOND TIME #27 FAILED, WHICH IS WHY RESTORATION DATE PRECEDES ABUTMENT DATE.
Patient problems: Inadequate Osseointegration
MAUDE report 15602934 ↗InjuryJun 3, 2022Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PSYCHOLOGICAL DISORDER, BRUXISM, PAIN.
Patient problems: Inadequate Osseointegration
MAUDE report 14588065 ↗InjuryOct 31, 2021Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED - IMPLANT WAS PLACED RIGHT AFTER EXPLANTATION OF AN OLDER IMPLANT - IMPLANT DIAMETER WAS TOO BIG.
Patient problems: Inadequate Osseointegration
MAUDE report 12725534 ↗InjuryJun 22, 2021Failure to Osseointegrate; Failure to Osseointegrate
InjuryFeb 5, 2021Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 11285738 ↗InjuryAug 10, 2020Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED. 6.0 IMPLANT PLACED AND WAS TOO WIDE, CHANGED TO 5.0 BUT WAS GOING TO BE TOO LONG AND PLACED 5.0X9.5L. SAME PATIENT AS (B)(6).
Patient problems: No Code Available
MAUDE report 10383685 ↗InjuryFeb 25, 2019Failure to Osseointegrate; Failure to Osseointegrate
NO OSSEOINTEGRATION, THREAD TAP USED, IMPLANT WAS COMPLETELY COVERED BY BONE, IMMEDIATE IMPLANTATION, INFECTION, MOBILITY CHECK-UP AND LASER DONE THE SAME DAY.
Patient problems: No Code Available
MAUDE report 8365913 ↗
Other Implant, Endosseous, Root-Form devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page