Skip to content

SPI ELEMENT INICELL 4.23.532Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI ELEMENT INICELL (67)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Thommen SPI Element Inicell implant is a root-form implant placed in bone. All 172 reports are filed as injuries, and most describe the implant failing to bond with the bone or losing that bond, with some noting infection, or an implant that was not firmly held at placement. The reports mention no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
173
Deaths reported
0
Injuries reported
173
Recalls naming it
0

Summary

The FDA's MAUDE database contains 173 adverse event reports that list this exact device (DI 07640156470130), received between Feb 15, 2018 and Sep 28, 2026 — classified as 173 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (139 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 2 reportsJun 2024: 3 reportsJul 2024: 9 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 4 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 3 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 12 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate139
  • Loss of Osseointegration32
  • Break1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Inadequate Osseointegration126
  • No Code Available45
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2

Recent adverse event reports

The 25 most recent of 173 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Break

    Implant fracture.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 50911592 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    Patient problems: Inadequate Osseointegration

    MAUDE report 50473390 ↗
  • InjuryMay 19, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERI-IMPLANTITIS, INFECTION, PAIN, BLEEDING, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25214940 ↗
  • InjuryMay 19, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PAIN, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25214937 ↗
  • InjuryMay 19, 2026Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25214844 ↗
  • InjuryMar 6, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24529336 ↗
  • InjuryJan 30, 2026Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, SMOKER.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24221763 ↗
  • InjuryJan 30, 2026Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, SMOKER.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24220421 ↗
  • InjuryJan 29, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24210282 ↗
  • InjuryJan 27, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24187869 ↗
  • InjuryJan 27, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24187842 ↗
  • InjuryJan 23, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24158870 ↗
  • InjuryJan 12, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24036244 ↗
  • InjuryJan 12, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24035752 ↗
  • InjuryJan 9, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING (B)(6) ARE SAME PATIENT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24019578 ↗
  • InjuryJan 9, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24018173 ↗
  • InjuryJan 8, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24001860 ↗
  • InjuryJan 7, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 23988587 ↗
  • InjuryNov 17, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, PERI-IMPLANTITIS, INFECTION, NUMBNESS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23567650 ↗
  • InjuryNov 10, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, BONE RESORPTION, INFECTION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23508624 ↗
  • InjuryOct 29, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 23413186 ↗
  • InjuryAug 26, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22892609 ↗
  • InjuryAug 4, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22698608 ↗
  • InjuryJul 1, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PRECEDING/SIMULTANEOUS BONE AUGMENTATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22368893 ↗
  • InjuryJun 18, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, XEROSTOMIA, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22283916 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page