Skip to content

SPI ELEMENT INICELL 4.23.723Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI ELEMENT INICELL (67)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Thommen Medical SPI ELEMENT INICELL dental implant, a screw-shaped titanium implant placed into the jawbone to replace a tooth. The reports mostly describe the implant failing to bond with the surrounding bone, or losing that bond after surgery, sometimes with mobility, swelling, infection or a fistula, and several mention smoking, diabetes or gum disease. There are 34 reports, all describing injuries and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
34
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 07640156470208), received between Jul 28, 2018 and Jul 14, 2026 — classified as 34 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (28 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate28
  • Loss of Osseointegration6

Patient problems

  • Inadequate Osseointegration27
  • No Code Available7

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 14, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25809977 ↗
  • InjuryMar 6, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24528578 ↗
  • InjuryAug 5, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERIODONTAL DISEASE, INFECTION, ABSCESS, INFLAMMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22712600 ↗
  • InjuryApr 4, 2025Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED

    Patient problems: Inadequate Osseointegration

    MAUDE report 21771314 ↗
  • InjuryFeb 13, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, DIABETES MELLITUS, SMOKER.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21373793 ↗
  • InjuryFeb 6, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION, PAIN, INCREASED SENSITIVITY, SWELLING, FISTULA, INFLAMMATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21319891 ↗
  • InjuryOct 14, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, ABSCESS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20438400 ↗
  • InjurySep 20, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20264237 ↗
  • InjurySep 13, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20211202 ↗
  • InjuryJun 24, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19593698 ↗
  • InjuryJun 10, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION (2024_1650 AND 2024_1651 ARE SAME PATIENT)

    Patient problems: Inadequate Osseointegration

    MAUDE report 19496402 ↗
  • InjuryJun 10, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION (2024_1650 AND 2024_1651 ARE SAME PATIENT)

    Patient problems: Inadequate Osseointegration

    MAUDE report 19496397 ↗
  • InjuryApr 23, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, NO THREAD TAP USED, INADEQUATE BONE QUALITY/QUANTITY - ANOTHER SIZE OF IMPLANT WAS CHOSEN.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19161610 ↗
  • InjuryFeb 17, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INFECTION, PAIN, SWELLING, INFLAMMATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18728190 ↗
  • InjuryJan 19, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, SMOKER, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18541280 ↗
  • InjuryOct 9, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT LOSS DUE TO CHANGE OF COLLAR HEIGHT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17894038 ↗
  • InjuryJul 20, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED. ANOTHER IMPLANT SIZE HAS BEEN PLACED SUCCESSFULLY. (B)(6) ARE THE SAME PATIENT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17359774 ↗
  • InjuryJan 21, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, PAIN, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16220897 ↗
  • InjuryJan 20, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INFLAMMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16214815 ↗
  • InjuryMay 27, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, SMOKER, BIOMECHANICAL OVERLOAD.

    Patient problems: Inadequate Osseointegration

    MAUDE report 14521037 ↗
  • InjuryFeb 23, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, COMPLICATION IN SITE PREPARATION, IMMEDIATE IMPLANTATION, INFECTION, INFLAMMATION ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 13583125 ↗
  • InjuryJan 26, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERI-IMPLANTITIS, INFECTION, PAIN, INFLAMMATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 13354962 ↗
  • InjuryOct 22, 2021Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, ILLNESS REQUIRING STEROIDS, PSYCHOLOGICAL DISORDER, XEROSTOMIA, PAIN, SWELLING. PATIENT PRESENTED WITH IMPLANT #6 OUT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 12679235 ↗
  • InjuryJul 4, 2021Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, SMOKER, ANOTHER IMPLANT SIZE WAS CHOSEN ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 12114184 ↗
  • InjuryJun 19, 2021Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PAIN, MOBILITY ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 12029926 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page