Skip to content

SPI CONTACT INICELL 4.23.626Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI CONTACT INICELL (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a SPI Contact Inicell dental implant by Thommen Medical AG, a root-form screw implant placed in the jawbone. The 27 reports mostly describe the implant failing to fuse with the surrounding bone, sometimes after it had already gained initial stability, with reports also mentioning swelling, bleeding, pain, infection and implant mobility. All 27 reports describe injuries, with no deaths. No recalls name this device, and report numbers are flat at 2 in each of the last two 12-month periods.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
27
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 07640156470611), received between Mar 11, 2020 and Jan 29, 2026 — classified as 27 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (25 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate25
  • Loss of Osseointegration2

Patient problems

  • Inadequate Osseointegration24
  • No Code Available3

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 29, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED

    Patient problems: Inadequate Osseointegration

    MAUDE report 24207978 ↗
  • InjuryJan 22, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, SMOKER, PERIODONTAL DISEASE, COMPLICATION IN SITE PREPARATION, PERI-IMPLANTITIS, INFECTION, PAIN, ABSCESS, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24142496 ↗
  • InjuryAug 20, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SWELLING, BLEEDING, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22849289 ↗
  • InjuryMar 19, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION, SWELLING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21642157 ↗
  • InjuryJul 1, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, SWELLING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19644338 ↗
  • InjuryApr 5, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY, VESTIBULAR BONE WALL UNSTABLE, WILL BE RE-IMPLANTED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19048769 ↗
  • InjuryFeb 1, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, SMOKER, INADEQUATE BONE QUALITY/QUANTITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18620557 ↗
  • InjuryJan 28, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, NO THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, BRUXISM, SWELLING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18592879 ↗
  • InjuryAug 25, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PAIN. (B)(6) ARE THE SAME PATIENT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17622224 ↗
  • InjuryJul 26, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, DISEASED MUCOUS MEMBRANE, IMMEDIATE IMPLANTATION, PAIN, MOBILITY, IMMEDIATE LOAD. (B)(6) ARE THE SAME PATIENT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17394890 ↗
  • InjuryJun 7, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, COMPLICATION IN SITE PREPARATION, IMMEDIATE IMPLANTATION, PAIN, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17078896 ↗
  • InjuryJan 24, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERIODONTAL DISEASE, COMPLICATION IN SITE PREPARATION, IMMEDIATE IMPLANTATION - SOCKET TOO LARGE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16231819 ↗
  • InjuryDec 29, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, IMMEDIATE IMPLANTATION, PRECEDING/SIMULTANEOUS BONE AUGMENTATION, INFECTION, FISTULA, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16066156 ↗
  • InjuryOct 14, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION, PAIN, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15602712 ↗
  • InjurySep 29, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15507670 ↗
  • InjurySep 22, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, BRUXISM, IMMEDIATE IMPLANTATION, PRECEDING/SIMULTANEOUS BONE AUGMENTATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15469278 ↗
  • InjurySep 22, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, NO THREAD TAP USED, IMMEDIATE IMPLANTATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15469157 ↗
  • InjurySep 6, 2022Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15360487 ↗
  • InjurySep 6, 2022Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION (B)(6) AND (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 15360438 ↗
  • InjuryAug 29, 2022Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15310361 ↗
  • InjuryAug 29, 2022Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMMEDIATE IMPLANTATION. IMPLANT DID NOT TAP / SOCKET TOO LARGE. TOO LITTLE INSERTION TORQUE. PLACED/REMOVED THE SAME DAY IN EXTRACTION SOCKET.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15310170 ↗
  • InjuryJul 8, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 14953904 ↗
  • InjuryDec 21, 2021Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 13050683 ↗
  • InjurySep 6, 2021Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 12427986 ↗
  • InjuryNov 16, 2020Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, COMPROMISED IMMUNE RESISTANCE, SMOKER, PERI-IMPLANTITIS, INFECTION, SWELLING, ABSCESS, MOBILITY.

    Patient problems: No Code Available

    MAUDE report 10842134 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page