Quick summary

This is a Thommen Medical dental implant, a screw-shaped root-form implant used to replace missing teeth. The 64 reports mostly describe the implant failing to fuse with the surrounding bone, or losing that fusion over time, which can leave it loose, sometimes alongside infection, sinus perforation, or gum disease. All 64 reports describe injuries, with no deaths. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
64
Deaths reported
0
Injuries reported
64
Recalls naming it
0

Summary

The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 07640156470840), received between Apr 27, 2018 and Jan 29, 2026 — classified as 64 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (49 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 4 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 7 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate49
  • Loss of Osseointegration15

Patient problems

  • Inadequate Osseointegration45
  • No Code Available19

Recent adverse event reports

The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 29, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, PERIODONTAL DISEASE, SINUS PERFORATION, INFECTION, ABSCESS, INFLAMMATION, AND MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24209710 ↗
  • InjuryJun 12, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, BONE RESORPTION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22191830 ↗
  • InjuryMay 12, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22003272 ↗
  • InjuryNov 28, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, NO THREAD TAP USED, COMPLICATION IN SITE PREPARATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20805578 ↗
  • InjuryNov 28, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE (B)(6) ARE SAME PATIENT)

    Patient problems: Inadequate Osseointegration

    MAUDE report 20804832 ↗
  • InjuryNov 19, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, BLOOD COAGULATION DISORDER, DIABETES MELLITUS, SMOKER.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20721481 ↗
  • InjuryNov 19, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, BLOOD COAGULATION DISORDER.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20721360 ↗
  • InjuryNov 12, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20663708 ↗
  • InjuryNov 7, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, DIABETES MELLITUS, PERI-IMPLANTITIS, PAIN, FISTULA, ABSCESS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20629093 ↗
  • InjuryNov 7, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED

    Patient problems: Inadequate Osseointegration

    MAUDE report 20628964 ↗
  • InjuryOct 24, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERI-IMPLANTITIS, PAIN, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20521368 ↗
  • InjuryOct 23, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SWELLING, MOBILITY

    Patient problems: Inadequate Osseointegration

    MAUDE report 20512751 ↗
  • InjuryJun 4, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19458027 ↗
  • InjuryJun 4, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19457731 ↗
  • InjuryMay 23, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, PERI-IMPLANTITIS, SWELLING, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19379068 ↗
  • InjuryMay 23, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19378956 ↗
  • InjuryMay 22, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, NO THREAD TAP USED ((B)(6) AND (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 19371994 ↗
  • InjuryMay 22, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, NO THREAD TAP USED, COMPLICATION IN SITE PREPARATION (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 19368893 ↗
  • InjuryDec 18, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18344574 ↗
  • InjuryNov 20, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, SMOKER, PERIODONTAL DISEASE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18169587 ↗
  • InjuryNov 14, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18128594 ↗
  • InjuryNov 14, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, NO THREAD TAP USED, COMPLICATION IN SITE PREPARATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18128492 ↗
  • InjuryJul 13, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, NUMBNESS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17313891 ↗
  • InjuryJul 12, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION, INFECTION, INFLAMMATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17306357 ↗
  • InjuryJun 29, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, PERI-IMPLANTITIS, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17223048 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →