Skip to content

SPI ELEMENT INICELL 4.23.131Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI ELEMENT INICELL (67)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Thommen SPI Element Inicell implant is placed in bone. The 284 reports mostly describe implants that did not fuse with the bone or later lost their hold. All 284 are injury reports, with no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
284
Deaths reported
0
Injuries reported
284
Recalls naming it
0

Summary

The FDA's MAUDE database contains 284 adverse event reports that list this exact device (DI 07640156470871), received between Feb 15, 2018 and Jul 10, 2026 — classified as 284 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (218 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 5 reportsJan 2024: 1 reportsFeb 2024: 12 reportsMar 2024: 3 reportsApr 2024: 6 reportsMay 2024: 6 reportsJun 2024: 10 reportsJul 2024: 10 reportsAug 2024: 4 reportsSep 2024: 8 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 6 reportsFeb 2025: 2 reportsMar 2025: 3 reportsApr 2025: 1 reportsMay 2025: 9 reportsJun 2025: 0 reportsJul 2025: 4 reportsAug 2025: 1 reportsSep 2025: 6 reportsOct 2025: 2 reportsNov 2025: 7 reportsDec 2025: 0 reportsJan 2026: 4 reportsFeb 2026: 2 reportsMar 2026: 6 reportsApr 2026: 4 reportsMay 2026: 1 reportsJun 2026: 3 reportsJul 2026: 3 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate218
  • Loss of Osseointegration66

Patient problems

  • Inadequate Osseointegration221
  • No Code Available63

Recent adverse event reports

The 25 most recent of 284 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 10, 2026Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25776193 ↗
  • InjuryJul 7, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25733525 ↗
  • InjuryJul 7, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PAIN (B)(6) ARE SAME PATIENT)

    Patient problems: Inadequate Osseointegration

    MAUDE report 25730145 ↗
  • InjuryJun 22, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25577136 ↗
  • InjuryJun 16, 2026Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25516593 ↗
  • InjuryJun 15, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, COMPLICATION IN SITE PREPARATION, IMMEDIATE IMPLANTATION, OVERHEATING OF BONE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25500907 ↗
  • InjuryMay 26, 2026Failure to Osseointegrate

    IMPLANT EXPLANTATION - NERVE ENCROACHMENT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25275436 ↗
  • InjuryApr 28, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, PERI-IMPLANTITIS, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25016813 ↗
  • InjuryApr 21, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, INFLAMMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24937897 ↗
  • InjuryApr 8, 2026Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, COMPLICATION IN SITE PREPARATION, INADEQUATE BONE QUALITY/QUANTITY (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24813012 ↗
  • InjuryApr 7, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, OVERHEATING OF BONE, INFLAMMATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24800939 ↗
  • InjuryMar 26, 2026Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24698528 ↗
  • InjuryMar 26, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, OVERHEATING OF BONE, INFLAMMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24697829 ↗
  • InjuryMar 6, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, ABSCESS, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24529368 ↗
  • InjuryMar 6, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24528555 ↗
  • InjuryMar 5, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PERI-IMPLANTITIS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24516751 ↗
  • InjuryMar 4, 2026Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INADEQUATE BONE QUALITY/QUANTITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24505038 ↗
  • InjuryFeb 27, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24467517 ↗
  • InjuryFeb 19, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERI-IMPLANTITIS, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24389818 ↗
  • InjuryJan 27, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24188277 ↗
  • InjuryJan 13, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24051289 ↗
  • InjuryJan 9, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24019112 ↗
  • InjuryJan 9, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24019107 ↗
  • InjuryNov 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, SWELLING, FISTULA, INFLAMMATION ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 23669768 ↗
  • InjuryNov 27, 2025Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23664724 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page