Skip to content

SPI ELEMENT INICELL 4.23.121Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI ELEMENT INICELL (67)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Thommen SPI Element dental implant is placed in the jawbone to hold a replacement tooth. All 351 reports are injuries with no deaths, and most describe the implant failing to fuse with the bone or losing its fusion. Some reports mention poor bone quality or the implant being loose. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
349
Deaths reported
0
Injuries reported
349
Recalls naming it
0

Summary

The FDA's MAUDE database contains 349 adverse event reports that list this exact device (DI 07640156470888), received between Feb 16, 2018 and Jul 8, 2026 — classified as 349 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (247 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 2 reportsJan 2024: 7 reportsFeb 2024: 5 reportsMar 2024: 7 reportsApr 2024: 4 reportsMay 2024: 3 reportsJun 2024: 4 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 16 reportsOct 2024: 2 reportsNov 2024: 4 reportsDec 2024: 2 reportsJan 2025: 6 reportsFeb 2025: 4 reportsMar 2025: 3 reportsApr 2025: 1 reportsMay 2025: 3 reportsJun 2025: 4 reportsJul 2025: 0 reportsAug 2025: 7 reportsSep 2025: 3 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 6 reportsFeb 2026: 6 reportsMar 2026: 3 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 5 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate247
  • Loss of Osseointegration101
  • Break1

Patient problems

  • Inadequate Osseointegration240
  • No Code Available108

Recent adverse event reports

The 25 most recent of 349 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 8, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY

    Patient problems: Inadequate Osseointegration

    MAUDE report 25743865 ↗
  • InjuryJul 7, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION (B)(6) ARE SAME PATIENT)

    Patient problems: Inadequate Osseointegration

    MAUDE report 25728930 ↗
  • InjuryJul 7, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25728845 ↗
  • InjuryJul 6, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25713681 ↗
  • InjuryJul 6, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25713669 ↗
  • InjuryApr 29, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25030716 ↗
  • InjuryMar 15, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, INCREASED SENSITIVITY, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24601580 ↗
  • InjuryMar 6, 2026Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, NO THREAD TAP USED ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24528485 ↗
  • InjuryMar 4, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, NO THREAD TAP USED, INCREASED SENSITIVITY, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24505016 ↗
  • InjuryFeb 27, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PERI-IMPLANTITIS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24468195 ↗
  • InjuryFeb 23, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, BLEEDING, INFLAMMATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24412030 ↗
  • InjuryFeb 5, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, BONE RESORPTION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24265805 ↗
  • InjuryFeb 5, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24265706 ↗
  • InjuryFeb 5, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, PAIN, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24265687 ↗
  • InjuryFeb 4, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERI-IMPLANTITIS, SWELLING, BLEEDING (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24256681 ↗
  • InjuryJan 23, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24158685 ↗
  • InjuryJan 19, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24107182 ↗
  • InjuryJan 9, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 24019561 ↗
  • InjuryJan 9, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24017882 ↗
  • InjuryJan 7, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 23988536 ↗
  • InjuryJan 7, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, MOBILITY (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 23987672 ↗
  • InjuryNov 24, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, BLEEDING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23631158 ↗
  • InjuryNov 19, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PAIN.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23593111 ↗
  • InjuryOct 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23402708 ↗
  • InjurySep 2, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22937753 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page