SPI CONTACT INICELL 4.23.254Recalls & reported problems
Thommen Medical AG · Class 2 · Implant
All models of SPI CONTACT INICELL (37)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 20
- Deaths reported
- 0
- Injuries reported
- 20
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 07640156470918), received between Mar 16, 2018 and Sep 17, 2025 — classified as 20 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate13
- Loss of Osseointegration7
Patient problems
- Inadequate Osseointegration11
- No Code Available9
Recent adverse event reports
The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.
InjurySep 17, 2025Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED (B)(6) ARE SAME PATIENT.
Patient problems: Inadequate Osseointegration
MAUDE report 23074016 ↗InjuryAug 20, 2025Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, BLEEDING, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 22850318 ↗InjuryMay 1, 2025Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 21941084 ↗InjurySep 15, 2023Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.
Patient problems: Inadequate Osseointegration
MAUDE report 17754265 ↗InjuryMay 24, 2023Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.
Patient problems: Inadequate Osseointegration
MAUDE report 16991010 ↗InjuryMay 8, 2023Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, THREAD TAP USED, DIABETES MELLITUS.
Patient problems: Inadequate Osseointegration
MAUDE report 16884770 ↗InjuryAug 29, 2022Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, THREAD TAP USED.
Patient problems: Inadequate Osseointegration
MAUDE report 15309222 ↗InjuryJun 30, 2021Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 12091921 ↗InjuryJun 23, 2021Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 12047730 ↗InjuryJun 22, 2021Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, SMOKER, PERI-IMPLANTITIS, PAIN, FISTULA, INFLAMMATION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 12041136 ↗InjuryFeb 26, 2021Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 11383953 ↗InjuryOct 30, 2020Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY
Patient problems: No Code Available
MAUDE report 10761510 ↗InjuryJul 20, 2020Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: No Code Available
MAUDE report 10294762 ↗InjuryJan 31, 2020Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: No Code Available
MAUDE report 9652428 ↗InjuryDec 6, 2019Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: No Code Available
MAUDE report 9422617 ↗InjuryDec 6, 2019Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: No Code Available
MAUDE report 9422609 ↗InjuryJun 7, 2019Failure to Osseointegrate; Failure to Osseointegrate
NO OSSEOINTEGRATION, NO THREAD TAP USED, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED BY BONE, SENSITIVITY.
Patient problems: No Code Available
MAUDE report 8677733 ↗InjuryFeb 22, 2019Failure to Osseointegrate; Failure to Osseointegrate
NO PRIMARY STABILITY, NO THREAD TAP USED, IMPLANT WASN'T COMPLETELY COVERED BY BONE.
Patient problems: No Code Available
MAUDE report 8360831 ↗InjuryMar 20, 2018Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY ACHIEVED, INSTABILITY, NO OSSEOINTEGRATION.
Patient problems: No Code Available
MAUDE report 7353005 ↗InjuryMar 16, 2018Loss of Osseointegration; Loss of Osseointegration
PRIMARY STABILITY ACHIEVED, LOSS OF OSSEOINTEGRATION, INSTABILITY.
Patient problems: No Code Available
MAUDE report 7345394 ↗
Other Implant, Endosseous, Root-Form devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.