This Thommen SPI CONTACT INICELL dental implant has 77 adverse event reports naming its identifier, all describing injuries, with 77 injuries and no deaths: implants that failed to osseointegrate or lost osseointegration, sometimes with mobility and peri-implantitis. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 77
- Deaths reported
- 0
- Injuries reported
- 77
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 77 adverse event reports that list this exact device (DI 07640156470970), received between Feb 15, 2018 and May 19, 2026 — classified as 77 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (63 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate63
- Loss of Osseointegration14
Patient problems
- Inadequate Osseointegration44
- No Code Available33
Recent adverse event reports
The 25 most recent of 77 reports, in the reporter's own words as published by the FDA.
InjuryMay 19, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 25214874 ↗InjuryMay 19, 2026Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED (B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 25214817 ↗InjuryApr 21, 2026Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, PERI-IMPLANTITIS, MOBILITY (B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 24937434 ↗InjuryApr 21, 2026Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, PERI-IMPLANTITIS, MOBILITY (B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 24937429 ↗InjuryMar 26, 2026Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, SMOKER.
Patient problems: Inadequate Osseointegration
MAUDE report 24697662 ↗InjuryJan 30, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION (B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 24219604 ↗InjuryJan 26, 2026Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 24176432 ↗InjurySep 19, 2025Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 23093497 ↗InjurySep 17, 2025Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 23074490 ↗InjuryJun 17, 2025Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PAIN ((B)(6) AND (B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 22217504 ↗InjuryJun 16, 2025Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, INADEQUATE BONE QUALITY/QUANTITY.
Patient problems: Inadequate Osseointegration
MAUDE report 22209617 ↗InjuryJun 12, 2025Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY ((B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 22194337 ↗InjuryJun 12, 2025Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, INADEQUATE BONE QUALITY/QUANTITY ((B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 22194066 ↗InjuryMay 30, 2025Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 22113073 ↗InjuryMay 2, 2025Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, INCREASED SENSITIVITY, INFLAMMATION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 21948266 ↗InjuryMar 4, 2025Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY
Patient problems: Inadequate Osseointegration
MAUDE report 21512999 ↗InjuryFeb 13, 2025Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PRECEDING/SIMULTANEOUS BONE AUGMENTATION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 21372216 ↗InjuryFeb 6, 2025Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, SMOKER, IMMEDIATE IMPLANTATION.
Patient problems: Inadequate Osseointegration
MAUDE report 21319962 ↗InjurySep 30, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION.
Patient problems: Inadequate Osseointegration
MAUDE report 20331536 ↗InjuryJul 12, 2024Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, BRUXISM, MOBILITY
Patient problems: Inadequate Osseointegration
MAUDE report 19730803 ↗InjuryJul 2, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION ((B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 19653825 ↗InjuryApr 28, 2024Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION, INFECTION, PAIN, INFLAMMATION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 19199020 ↗InjuryApr 26, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.
Patient problems: Inadequate Osseointegration
MAUDE report 19189605 ↗InjuryApr 26, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.
Patient problems: Inadequate Osseointegration
MAUDE report 19189581 ↗InjuryNov 24, 2023Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, COMPLICATION IN SITE PREPARATION (SOFT BONE), MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 18203392 ↗
Other Implant, Endosseous, Root-Form devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page