Skip to content

SPI CONTACT INICELL 4.23.224Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI CONTACT INICELL (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Thommen SPI Contact dental implant has 106 reports, no deaths and 106 injuries. Most reports describe the implant failing to bond with the jawbone or losing that bond, with many noting poor bone quality or the implant not being fully covered with bone, and some describe the implant becoming loose. Reports fell from 16 in the prior 12 months to 2 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
106
Deaths reported
0
Injuries reported
106
Recalls naming it
0

Summary

The FDA's MAUDE database contains 106 adverse event reports that list this exact device (DI 07640156471052), received between Feb 24, 2018 and Aug 28, 2026 — classified as 106 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (83 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 3 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 4 reportsMar 2025: 1 reportsApr 2025: 3 reportsMay 2025: 5 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate83
  • Loss of Osseointegration23

Patient problems

  • Inadequate Osseointegration58
  • No Code Available48

Recent adverse event reports

The 25 most recent of 106 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 28, 2026Failure to Osseointegrate

    Implant didn't achieve osseointegration, implant wasn't completely covered with bone, no thread tap used, swelling, mobility.

    Patient problems: Inadequate Osseointegration

    MAUDE report 50492665 ↗
  • InjuryDec 30, 2025Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, INADEQUATE BONE QUALITY/QUANTITY, BONE RESORPTION ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 23921123 ↗
  • InjuryAug 5, 2025Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, INADEQUATE BONE QUALITY/QUANTITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22715506 ↗
  • InjuryJul 2, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22382743 ↗
  • InjuryJun 12, 2025Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22194245 ↗
  • InjuryMay 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22095375 ↗
  • InjuryMay 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, MOBILITY ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 22095352 ↗
  • InjuryMay 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22095275 ↗
  • InjuryMay 28, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, INFLAMMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22095233 ↗
  • InjuryMay 27, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PAIN ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 22086507 ↗
  • InjuryApr 22, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, IMMEDIATE IMPLANTATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21879905 ↗
  • InjuryApr 1, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION, INCREASED SENSITIVITY (B)(6), ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 21739283 ↗
  • InjuryApr 1, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION, INCREASED SENSITIVITY (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 21739272 ↗
  • InjuryMar 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21719045 ↗
  • InjuryFeb 13, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PAIN.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21372978 ↗
  • InjuryFeb 5, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21310694 ↗
  • InjuryFeb 5, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, PERI-IMPLANTITIS, PAIN, INCREASED SENSITIVITY, MOBILITY - BRIDGE WITH 5 VARIOMULTI ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 21310544 ↗
  • InjuryFeb 5, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, PERI-IMPLANTITIS, PAIN, INCREASED SENSITIVITY, MOBILITY - BRIDGE WITH 5 VARIOMULTI ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 21310469 ↗
  • InjurySep 6, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20154744 ↗
  • InjuryJun 18, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PAIN (2024_1760 AND 2024_1761 ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 19557492 ↗
  • InjuryFeb 28, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18795226 ↗
  • InjuryFeb 16, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMMEDIATE IMPLANTATION, INADEQUATE BONE QUALITY/QUANTITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18721454 ↗
  • InjuryFeb 13, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERIODONTAL DISEASE, BIOMECHANICAL OVERLOAD, PRECEDING/SIMULTANEOUS BONE AUGMENTATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18693740 ↗
  • InjuryJan 3, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PAIN, SWELLING, BLEEDING, INFLAMMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18434055 ↗
  • InjuryNov 17, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18157967 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page