Skip to content

SPI ELEMENT INICELL 4.23.136Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI ELEMENT INICELL (67)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Thommen Medical dental implant used as a screw-like root replacement in the jaw. The 46 reports mostly describe the implant failing to fuse with the surrounding bone, or losing that bond later, sometimes with mobility, infection, abscess and pain; one report describes the implant itself breaking. All 46 reports involve injuries, and no deaths are reported. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
46
Deaths reported
0
Injuries reported
46
Recalls naming it
0

Summary

The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 07640156471168), received between Feb 2, 2018 and Jul 6, 2026 — classified as 46 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (36 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate36
  • Loss of Osseointegration8
  • Break1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Inadequate Osseointegration33
  • No Code Available12
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 6, 2026Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY ((B)(6) AND (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 25713934 ↗
  • InjuryJul 6, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERI-IMPLANTITIS, PAIN, ABSCESS, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25713923 ↗
  • InjuryJul 6, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERI-IMPLANTITIS, ABSCESS, INFLAMMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25713899 ↗
  • InjuryJan 27, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24188221 ↗
  • InjuryNov 24, 2025Break; Break

    IMPLANT FRACTURE, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INNER THREAD OF THE IMPLANT IS DEFECTIVE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23628798 ↗
  • InjuryNov 19, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMMEDIATE IMPLANTATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23592799 ↗
  • InjuryOct 23, 2025Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, COMPLICATION IN SITE PREPARATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23365412 ↗
  • InjuryAug 25, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, BRUXISM, IMMEDIATE IMPLANTATION, INADEQUATE BONE QUALITY/QUANTITY, INFLAMMATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22882345 ↗
  • InjuryMay 23, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22076258 ↗
  • InjuryFeb 11, 2025Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, BRUXISM.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21353876 ↗
  • InjuryDec 5, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, COMPLICATION IN SITE PREPARATION, PAIN, SWELLING, INFLAMMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20850744 ↗
  • InjuryNov 5, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, OVERHEATING OF BONE - IMMEDIATE LOADING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20607952 ↗
  • InjuryOct 31, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PAIN.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20576208 ↗
  • InjurySep 11, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, INCREASED SENSITIVITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20188063 ↗
  • InjurySep 10, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, SMOKER, IMMEDIATE IMPLANTATION, MOBILITY (B)(6) ARE SAME PATIENT)

    Patient problems: Inadequate Osseointegration

    MAUDE report 20179223 ↗
  • InjuryJun 28, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERI-IMPLANTITIS, INFECTION, PAIN, INFLAMMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19631086 ↗
  • InjuryJun 26, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, COMPLICATION IN SITE PREPARATION, IMMEDIATE IMPLANTATION, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 19614770 ↗
  • InjuryJun 26, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, COMPLICATION IN SITE PREPARATION, IMMEDIATE IMPLANTATION, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 19614730 ↗
  • InjuryMay 2, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERI-IMPLANTITIS, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19227919 ↗
  • InjuryMay 2, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, INFECTION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19227904 ↗
  • InjuryApr 18, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PAIN, MOBILITY

    Patient problems: Inadequate Osseointegration

    MAUDE report 19133150 ↗
  • InjuryJan 31, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, COMPLICATION IN SITE PREPARATION, BIOMECHANICAL OVERLOAD, IMMEDIATE IMPLANTATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18612630 ↗
  • InjuryOct 16, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, IMMEDIATE IMPLANTATION, INFECTION, SWELLING, INFLAMMATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17938918 ↗
  • InjuryAug 31, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, NUMBNESS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17659652 ↗
  • InjuryJun 30, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17235668 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates