SPI ELEMENT INICELL 4.23.132Recalls & reported problems
Thommen Medical AG · Class 2 · Implant
All models of SPI ELEMENT INICELL (67)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Thommen SPI Element dental implant is a screw placed in the jaw bone to support a replacement tooth. Its 729 reports mostly describe the implant not fusing with the bone, with some also noting looseness or gum infection around the implant. The reports mention no deaths and 729 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 719
- Deaths reported
- 0
- Injuries reported
- 719
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 719 adverse event reports that list this exact device (DI 07640156471205), received between Jan 22, 2018 and Sep 28, 2026 — classified as 719 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (558 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate558
- Loss of Osseointegration157
- Loosening of Implant Not Related to Bone-Ingrowth4
Patient problems
- Inadequate Osseointegration553
- No Code Available160
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
Recent adverse event reports
The 25 most recent of 719 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Failure to Osseointegrate
Primary stability not achieved, implant was completely covered with bone, no thread tap used - another size of implant was chosen.
Patient problems: Inadequate Osseointegration
MAUDE report 50911124 ↗InjuryAug 27, 2026Failure to Osseointegrate
Primary stability not achieved ((b)(6) are same patient).
Patient problems: Inadequate Osseointegration
MAUDE report 50469862 ↗InjuryAug 11, 2026Failure to Osseointegrate
Implant didn't achieve osseointegration, mobility.
Patient problems: Inadequate Osseointegration
MAUDE report 50207392 ↗InjuryAug 5, 2026Loss of Osseointegration
Loss of osseointegration, peri-implantitis.
Patient problems: Inadequate Osseointegration
MAUDE report 50141101 ↗InjuryJul 14, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 25809907 ↗InjuryJul 13, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED.
Patient problems: Inadequate Osseointegration
MAUDE report 25796340 ↗InjuryJul 13, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, INFLAMMATION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 25792591 ↗InjuryJul 10, 2026Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, PERIODONTAL DISEASE, BIOMECHANICAL OVERLOAD, PERI-IMPLANTITIS, PAIN, NUMBNESS, INCREASED SENSITIVITY, BLEEDING, INFLAMMATION
Patient problems: Inadequate Osseointegration
MAUDE report 25776102 ↗InjuryJul 10, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, SWELLING, MOBILITY
Patient problems: Inadequate Osseointegration
MAUDE report 25775791 ↗InjuryJul 8, 2026Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY
Patient problems: Inadequate Osseointegration
MAUDE report 25744262 ↗InjuryJun 22, 2026Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, PERI-IMPLANTITIS, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 25577202 ↗InjuryJun 22, 2026Failure to Osseointegrate
InjuryJun 22, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY ((B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 25576025 ↗InjuryJun 15, 2026Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, NO THREAD TAP USED ((B)(6) AND (B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 25501600 ↗InjuryJun 2, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PRECEDING/SIMULTANEOUS BONE AUGMENTATION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 25360902 ↗InjuryMay 27, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 25299338 ↗InjuryMay 26, 2026Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PRECEDING/SIMULTANEOUS BONE AUGMENTATION - ANOTHER SIZE OF IMPLANT WAS CHOSEN.
Patient problems: Inadequate Osseointegration
MAUDE report 25276009 ↗InjuryMay 26, 2026Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 25275858 ↗InjuryMay 19, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, BLOOD COAGULATION DISORDER.
Patient problems: Inadequate Osseointegration
MAUDE report 25214796 ↗InjuryApr 29, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, MOBILITY (B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 25025553 ↗InjuryApr 28, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, NO THREAD TAP USED, PAIN, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 25016218 ↗InjuryApr 27, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, THREAD TAP USED, PRECEDING/SIMULTANEOUS BONE AUGMENTATION.
Patient problems: Inadequate Osseointegration
MAUDE report 24997063 ↗InjuryApr 13, 2026Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE.
Patient problems: Inadequate Osseointegration
MAUDE report 24855899 ↗InjuryApr 10, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, IMMEDIATE IMPLANTATION, PAIN.
Patient problems: Inadequate Osseointegration
MAUDE report 24843600 ↗InjuryMar 26, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 24698373 ↗
Other Implant, Endosseous, Root-Form devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.