SPI ELEMENT 4.13.952Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI ELEMENT (28)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07640156471328), received between Feb 2, 2018 and Jan 22, 2026 — classified as 8 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate6
  • Loss of Osseointegration2

Patient problems

  • No Code Available5
  • Inadequate Osseointegration3

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 22, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, BIOMECHANICAL OVERLOAD, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24144168 ↗
  • InjuryAug 6, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PERIODONTAL DISEASE, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19916961 ↗
  • InjurySep 29, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15508493 ↗
  • InjuryJul 28, 2020Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, SMOKER, BIOMECHANICAL OVERLOAD, PAIN, INFLAMMATION, MOBILITY.

    Patient problems: No Code Available

    MAUDE report 10330999 ↗
  • InjuryMay 18, 2020Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, RADIATION TX-HEAD / NECK AREA, ILLNESS REQUIRING STEROIDS, IMMEDIATE IMPLANTATION, INADEQUATE BONE QUALITY/QUANTITY, PAIN, FISTULA, INFLAMMATION, MOBILITY ((B)(6) ARE SAME PATIENT).

    Patient problems: No Code Available

    MAUDE report 10064485 ↗
  • InjuryJan 2, 2020Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, NO THREAD TAP USED, MOBILITY.

    Patient problems: No Code Available

    MAUDE report 9537310 ↗
  • InjuryOct 6, 2018Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY ACHIEVED, NO OSSEOINTEGRATION. UNDERSIZED IMPLANT BED, PAIN, FISTULA.

    Patient problems: No Code Available

    MAUDE report 7941288 ↗
  • InjuryFeb 2, 2018Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY ACHIEVED, INSTABILITY, IMMEDIATE IMPLANTATION.

    Patient problems: No Code Available

    MAUDE report 7236610 ↗