Skip to content

SPI ELEMENT 4.13.951Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Thommen Medical AG root-form dental implant's 41 reports mostly describe loss of osseointegration or failure to osseointegrate, with two cases of break; there were 41 injuries and no deaths. There are no recalls. Reports dropped from 5 in the previous 12 months to 0 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
41
Deaths reported
0
Injuries reported
41
Recalls naming it
0

Summary

The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 07640156471335), received between Mar 10, 2018 and Oct 15, 2025 — classified as 41 injuries.

The most frequently reported specific device problem is “Loss of Osseointegration” (20 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of Osseointegration20
  • Failure to Osseointegrate19
  • Break2

Patient problems

  • Inadequate Osseointegration23
  • No Code Available16
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 15, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PERI-IMPLANTITIS (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 23299615 ↗
  • InjuryMay 21, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, SMOKER, IMMEDIATE IMPLANTATION, INFECTION, INFLAMMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22060095 ↗
  • InjuryFeb 24, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERI-IMPLANTITIS, INFLAMMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21445087 ↗
  • InjuryJan 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21250701 ↗
  • InjuryJan 7, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21089498 ↗
  • InjuryAug 23, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, SMOKER, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING. IMPLANTS #26 AND 27 PLACED AT THE SAME TIME ARE STILL OK.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20045510 ↗
  • InjuryAug 15, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19990053 ↗
  • InjuryJul 21, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19787895 ↗
  • InjuryApr 30, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, BLOOD COAGULATION DISORDER, PERIODONTAL DISEASE, IMMEDIATE IMPLANTATION, PERI-IMPLANTITIS ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 19212988 ↗
  • InjuryApr 12, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19094514 ↗
  • InjuryMar 18, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, SMOKER, PAIN, INFLAMMATION, MOBILITY ((B)(4) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 18920648 ↗
  • InjuryMar 18, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, SMOKER, PAIN, INFLAMMATION, MOBILITY ((B)(4) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 18920641 ↗
  • InjuryFeb 9, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMMEDIATE LOAD.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18674596 ↗
  • InjuryFeb 9, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18674503 ↗
  • InjuryOct 24, 2023Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, SMOKER, PAIN, INFLAMMATION, MOBILITY (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 17993775 ↗
  • InjuryOct 24, 2023Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, SMOKER, PAIN, INFLAMMATION, MOBILITY ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 17993773 ↗
  • InjurySep 15, 2023Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17754328 ↗
  • InjuryJun 14, 2023Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, NO THREAD TAP USED, SMOKER, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17123695 ↗
  • InjuryMay 3, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, COMPLICATION IN SITE PREPARATION, INADEQUATE BONE QUALITY/QUANTITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16860463 ↗
  • InjuryNov 2, 2022Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15713104 ↗
  • InjuryApr 22, 2022Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PRECEDING/SIMULTANEOUS BONE AUGMENTATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 14185113 ↗
  • InjuryMar 15, 2022Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, PERI-IMPLANTITIS.

    MAUDE report 13768509 ↗
  • InjurySep 7, 2021Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, INFECTION, PAIN, SWELLING, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 12432790 ↗
  • InjurySep 6, 2021Break; Break

    IMPLANT FRACTURE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12428271 ↗
  • InjuryMay 5, 2021Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, PERI-IMPLANTITIS, MOBILITY - FIRST CROWN WAS FRACTURED IN 2019 AND GOT REPLACED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 11776570 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates