Skip to content

SPI ELEMENT 4.13.921Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI ELEMENT (28)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07640156471441), received between May 19, 2018 and Oct 15, 2025 — classified as 10 injuries.

The most frequently reported specific device problem is “Loss of Osseointegration” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of Osseointegration7
  • Failure to Osseointegrate2
  • Break1

Patient problems

  • Inadequate Osseointegration6
  • No Code Available3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 15, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PERI-IMPLANTITIS (B)(6) ARE SAME PATIENT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23300008 ↗
  • InjuryOct 20, 2023Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, PERI-IMPLANTITIS, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17973412 ↗
  • InjuryMar 3, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16477098 ↗
  • InjuryJan 13, 2022Break; Break

    IMPLANT FRACTURE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13249560 ↗
  • InjuryJan 4, 2022Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, BIOMECHANICAL OVERLOAD DUE TO HYBRID PROSTHESIS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 13151154 ↗
  • InjurySep 28, 2021Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PERI-IMPLANTITIS ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 12540603 ↗
  • InjuryMar 12, 2021Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY. SAME PATIENT AS (B)(6).

    Patient problems: Inadequate Osseointegration

    MAUDE report 11471266 ↗
  • InjuryJan 2, 2020Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, UNCONTROLLED ENDOCRINE ILLNESS, SMOKER, INADEQUATE BONE QUALITY/QUANTITY, PRECEDING/SIMULTANEOUS BONE AUGMENTATION, INFECTION, SWELLING, FISTULA, INFLAMMATION, MOBILITY.

    Patient problems: No Code Available

    MAUDE report 9537306 ↗
  • InjuryNov 10, 2018Failure to Osseointegrate; Failure to Osseointegrate

    NO PRIMARY STABILITY ACHIEVED. IMMEDIATE IMPLANTATION.

    Patient problems: No Code Available

    MAUDE report 8057954 ↗
  • InjuryMay 19, 2018Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION. BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, MOBILITY.

    Patient problems: No Code Available

    MAUDE report 7528370 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page