SPI ELEMENT 4.13.921Recalls & reported problems
Thommen Medical AG · Class 2 · Implant
All models of SPI ELEMENT (28)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 07640156471441), received between May 19, 2018 and Oct 15, 2025 — classified as 10 injuries.
The most frequently reported specific device problem is “Loss of Osseointegration” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of Osseointegration7
- Failure to Osseointegrate2
- Break1
Patient problems
- Inadequate Osseointegration6
- No Code Available3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryOct 15, 2025Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, PERI-IMPLANTITIS (B)(6) ARE SAME PATIENT.
Patient problems: Inadequate Osseointegration
MAUDE report 23300008 ↗InjuryOct 20, 2023Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, PERI-IMPLANTITIS, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 17973412 ↗InjuryMar 3, 2023Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 16477098 ↗InjuryJan 13, 2022Break; Break
IMPLANT FRACTURE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13249560 ↗InjuryJan 4, 2022Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, BIOMECHANICAL OVERLOAD DUE TO HYBRID PROSTHESIS.
Patient problems: Inadequate Osseointegration
MAUDE report 13151154 ↗InjurySep 28, 2021Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, PERI-IMPLANTITIS ((B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 12540603 ↗InjuryMar 12, 2021Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY. SAME PATIENT AS (B)(6).
Patient problems: Inadequate Osseointegration
MAUDE report 11471266 ↗InjuryJan 2, 2020Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, UNCONTROLLED ENDOCRINE ILLNESS, SMOKER, INADEQUATE BONE QUALITY/QUANTITY, PRECEDING/SIMULTANEOUS BONE AUGMENTATION, INFECTION, SWELLING, FISTULA, INFLAMMATION, MOBILITY.
Patient problems: No Code Available
MAUDE report 9537306 ↗InjuryNov 10, 2018Failure to Osseointegrate; Failure to Osseointegrate
NO PRIMARY STABILITY ACHIEVED. IMMEDIATE IMPLANTATION.
Patient problems: No Code Available
MAUDE report 8057954 ↗InjuryMay 19, 2018Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION. BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, MOBILITY.
Patient problems: No Code Available
MAUDE report 7528370 ↗
Other Implant, Endosseous, Root-Form devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026