This is a Thomomen Medical endosseous, root-form dental implant used as a screw-like anchor in the jawbone. The 35 reports mostly describe the implant failing to fuse with the surrounding bone, or losing that bond after it had formed, so the implant becomes loose or mobile; some also mention infection, pain, bleeding, and swelling, often alongside factors like poor bone quality, smoking, or diabetes. All 35 reports involve injuries, with no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 35
- Deaths reported
- 0
- Injuries reported
- 35
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 07640156471472), received between Mar 1, 2018 and May 23, 2025 — classified as 35 injuries.
The most frequently reported specific device problem is “Loss of Osseointegration” (18 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of Osseointegration18
- Failure to Osseointegrate17
Patient problems
- Inadequate Osseointegration22
- No Code Available13
Recent adverse event reports
The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.
InjuryMay 23, 2025Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, INADEQUATE BONE QUALITY/QUANTITY
Patient problems: Inadequate Osseointegration
MAUDE report 22077049 ↗InjuryOct 25, 2024Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PAIN, BLEEDING, INFLAMMATION, MOBILITY
Patient problems: Inadequate Osseointegration
MAUDE report 20530921 ↗InjuryAug 15, 2024Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, MOBILITY
Patient problems: Inadequate Osseointegration
MAUDE report 19989504 ↗InjuryAug 3, 2024Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, INFECTION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 19901716 ↗InjuryJul 21, 2024Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 19787894 ↗InjuryJul 16, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, SMOKER, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 19748362 ↗InjuryJul 11, 2024Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, RADIATION TX-HEAD / NECK AREA, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 19717953 ↗InjuryMar 18, 2024Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, INADEQUATE BONE QUALITY/QUANTITY.
Patient problems: Inadequate Osseointegration
MAUDE report 18919773 ↗InjurySep 21, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED.
Patient problems: Inadequate Osseointegration
MAUDE report 17787628 ↗InjuryAug 31, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, BONE RESORPTION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 17659838 ↗InjuryJun 15, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, INFECTION, FISTULA.
Patient problems: Inadequate Osseointegration
MAUDE report 17132434 ↗InjuryJun 7, 2023Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, PERI-IMPLANTITIS, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 17077841 ↗InjuryJun 1, 2023Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, PERI-IMPLANTITIS, INFLAMMATION.
Patient problems: Inadequate Osseointegration
MAUDE report 17037004 ↗InjuryApr 23, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED.
Patient problems: Inadequate Osseointegration
MAUDE report 16792775 ↗InjuryMar 3, 2023Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, INFECTION.
Patient problems: Inadequate Osseointegration
MAUDE report 16476727 ↗InjuryApr 25, 2022Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, BRUXISM, INCREASED SENSITIVITY, BLEEDING, INFLAMMATION.
Patient problems: Inadequate Osseointegration
MAUDE report 14192013 ↗InjuryJan 28, 2022Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, NO THREAD TAP USED, IMMEDIATE IMPLANTATION.
Patient problems: Inadequate Osseointegration
MAUDE report 13384673 ↗InjuryJan 26, 2022Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, FISTULA, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 13358021 ↗InjuryDec 17, 2021Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, DIABETES MELLITUS, PERI-IMPLANTITIS, PAIN, SEVERE BONE LOSS.
Patient problems: Inadequate Osseointegration
MAUDE report 13024862 ↗InjuryNov 8, 2021Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERIODONTAL DISEASE, PAIN, MOBILITY (2021_1944, 2021_1945 AND 2021_1946 ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 12766514 ↗InjuryNov 8, 2021Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERIODONTAL DISEASE, PAIN, MOBILITY (2021_1944, 2021_1945 AND 2021_1946 ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 12766509 ↗InjurySep 28, 2021Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, PERI-IMPLANTITIS ( (B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 12540617 ↗InjuryJan 8, 2020Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, FISTULA, BLEEDING, INFLAMMATION, MOBILITY.
Patient problems: No Code Available
MAUDE report 9562222 ↗InjuryOct 29, 2019Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, INFECTION, ABSCESS (NEARLY ALL MEDICAL RECORDS ARE CROSSED, MOST PROBABLY THE DENTIST DIDN'T UNDERSTAND THE LANGUAGE).
Patient problems: No Code Available
MAUDE report 9249330 ↗InjuryMay 20, 2019Failure to Osseointegrate; Failure to Osseointegrate
THREAD TAP USED, PRIMARY STABILITY WAS ACHIEVED, IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, MOBILITY.
Patient problems: No Code Available
MAUDE report 8622938 ↗
Other Implant, Endosseous, Root-Form devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page