This is a Thomenn Medical SPI dental implant, a screw-shaped post placed into the jawbone to replace a tooth. The 65 reports mostly describe the implant failing to fuse with the surrounding bone, sometimes loosening and becoming mobile, with some reports also mentioning pain, swelling and infection around the implant site. There are 65 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 65
- Deaths reported
- 0
- Injuries reported
- 65
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 65 adverse event reports that list this exact device (DI 07640156471717), received between Apr 28, 2018 and Oct 28, 2024 — classified as 65 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (50 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate50
- Loss of Osseointegration15
Patient problems
- Inadequate Osseointegration42
- No Code Available23
Recent adverse event reports
The 25 most recent of 65 reports, in the reporter's own words as published by the FDA.
InjuryOct 28, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, PERI-IMPLANTITIS, INFECTION
Patient problems: Inadequate Osseointegration
MAUDE report 20543650 ↗InjuryOct 23, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PAIN
Patient problems: Inadequate Osseointegration
MAUDE report 20512503 ↗InjuryOct 23, 2024Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, OVERHEATING OF BONE, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 20512479 ↗InjuryMay 7, 2024Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 19257217 ↗InjuryJun 20, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, IMMEDIATE IMPLANTATION.
Patient problems: Inadequate Osseointegration
MAUDE report 17160905 ↗InjuryJun 14, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, SMOKER, PERIODONTAL DISEASE, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 17124512 ↗InjuryJun 14, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, INADEQUATE BONE QUALITY/QUANTITY, INFECTION, PAIN, SWELLING, MOBILITY ((B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 17124053 ↗InjuryMay 4, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, COMPLICATION IN SITE PREPARATION, OVERHEATING OF BONE (2023_0724 AND 2023_0725 ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 16866921 ↗InjuryMay 4, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INFECTION, SWELLING, BLEEDING, INFLAMMATION (2023_0712 AND 2023_0713 ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 16866609 ↗InjuryMay 4, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, SMOKER, PERIODONTAL DISEASE, PAIN, MOBILITY (2023_0709, 2023_0710 AND 2023_0711 ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 16866594 ↗InjuryApr 28, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY ((B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 16831766 ↗InjuryApr 28, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, THREAD TAP USED, BONE RESORPTION.
Patient problems: Inadequate Osseointegration
MAUDE report 16829054 ↗InjuryApr 27, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INADEQUATE BONE QUALITY/QUANTITY, PAIN, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 16822435 ↗InjuryApr 27, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SWELLING.
Patient problems: Inadequate Osseointegration
MAUDE report 16822339 ↗InjuryApr 27, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, INFECTION, SWELLING, ABSCESS, BLEEDING, INFLAMMATION.
Patient problems: Inadequate Osseointegration
MAUDE report 16822310 ↗InjuryApr 25, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED.
Patient problems: Inadequate Osseointegration
MAUDE report 16808463 ↗InjuryApr 25, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, THREAD TAP USED, INFECTION, PAIN.
Patient problems: Inadequate Osseointegration
MAUDE report 16807186 ↗InjuryApr 25, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, THREAD TAP USED, INFECTION.
Patient problems: Inadequate Osseointegration
MAUDE report 16805337 ↗InjuryApr 25, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, THREAD TAP USED, PERI-IMPLANTITIS, PAIN.
Patient problems: Inadequate Osseointegration
MAUDE report 16805267 ↗InjuryApr 25, 2023Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, INFECTION, PAIN, ABSCESS, MOBILITY
Patient problems: Inadequate Osseointegration
MAUDE report 16805204 ↗InjuryNov 9, 2022Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, INFECTION, PAIN, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 15758408 ↗InjuryNov 9, 2022Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SWELLING.
Patient problems: Inadequate Osseointegration
MAUDE report 15758383 ↗InjuryNov 9, 2022Loss of Osseointegration; Loss of Osseointegration
InjuryOct 28, 2022Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERI-IMPLANTITIS, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 15691174 ↗InjuryOct 17, 2022Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SINUS PERFORATION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 15613544 ↗
Other Implant, Endosseous, Root-Form devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page