Skip to content

SPI CONTACT 4.13.211Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI CONTACT (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Thommen Medical SPI dental implant is a root-form screw implant placed into the jawbone. The 29 reports mostly describe the implant failing to fuse with the surrounding bone, or losing that fusion later, so it does not stay stable and can become mobile. All 29 reports involve injuries, with no deaths mentioned. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
29
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 07640156471748), received between May 31, 2018 and Mar 26, 2024 — classified as 29 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (22 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 3 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate22
  • Loss of Osseointegration7

Patient problems

  • Inadequate Osseointegration19
  • No Code Available10

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 26, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, INADEQUATE BONE QUALITY/QUANTITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18976961 ↗
  • InjuryFeb 28, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, INADEQUATE BONE QUALITY/QUANTITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18795350 ↗
  • InjuryFeb 28, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, THREAD TAP USED, PERIODONTAL DISEASE, INADEQUATE BONE QUALITY/QUANTITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18795277 ↗
  • InjuryFeb 28, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, INADEQUATE BONE QUALITY/QUANTITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18795198 ↗
  • InjuryNov 14, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18128666 ↗
  • InjuryJun 20, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17160874 ↗
  • InjuryJun 20, 2023Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17160200 ↗
  • InjuryMay 4, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16868162 ↗
  • InjuryMay 4, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, COMPLICATION IN SITE PREPARATION, OVERHEATING OF BONE ((B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 16866928 ↗
  • InjuryMay 4, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, SMOKER, PERIODONTAL DISEASE, PAIN, MOBILITY (2023_0709, 2023_0710 AND 2023_0711 ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 16866590 ↗
  • InjuryMay 4, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, COMPLICATION IN SITE PREPARATION, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16866575 ↗
  • InjuryMay 4, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, SMOKER, PERIODONTAL DISEASE, COMPLICATION IN SITE PREPARATION, INADEQUATE BONE QUALITY/QUANTITY, INFECTION, BLEEDING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16866317 ↗
  • InjuryApr 25, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, INFECTION, SWELLING, MOBILITY (B)(6) ARE SAME PATIENT).

    Patient problems: Inadequate Osseointegration

    MAUDE report 16807452 ↗
  • InjurySep 22, 2022Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, BIOMECHANICAL OVERLOAD.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15469071 ↗
  • InjuryJul 28, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15119366 ↗
  • InjuryFeb 23, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, PAIN, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 13582825 ↗
  • InjuryFeb 7, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 13469088 ↗
  • InjurySep 11, 2021Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 12457581 ↗
  • InjuryFeb 15, 2021Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED. SAME PATIENT AS (B)(6).

    Patient problems: Inadequate Osseointegration

    MAUDE report 11327156 ↗
  • InjuryJul 13, 2020Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.

    Patient problems: No Code Available

    MAUDE report 10263543 ↗
  • InjuryMar 25, 2019Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION - PRIMARY STABILITY WAS ACHIEVED, THREAD TAP USED, SMOKER, DIABETES, PERIODONTAL DISEASE AT THE TIME OF SURGERY, BONE RESORPTION, PERIIMPLANTITIS, SWELLING, ABSCESS, BLEEDING, INFLAMMATION ((B)(6) ARE SAME PATIENT).

    Patient problems: No Code Available

    MAUDE report 8447887 ↗
  • InjuryMar 25, 2019Loss of Osseointegration; Loss of Osseointegration

    SMOKER, PERIODONTAL DISEASE, THREAD TAP USED, PRIMARY STABILITY WAS ACHIEVED, IMPLANT ACHIEVED OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED BY BONE, BRUXISM, BONE RESORPTION, PERI-IMPLANTITIS, SWELLING, FISTULA, ABSCESS ((B)(6) ARE SAME PATIENT).

    Patient problems: No Code Available

    MAUDE report 8447881 ↗
  • InjuryMar 25, 2019Loss of Osseointegration; Loss of Osseointegration

    SMOKER, PERIODONTAL DISEASE, THREAD TAP USED, PRIMARY STABILITY WAS ACHIEVED, IMPLANT ACHIEVED OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED BY BONE, BRUXISM, BONE RESORPTION, PERI-IMPLANTITIS, SWELLING, FISTULA, ABSCESS (2019_0473, 0474 AND 0475 ARE SAME PATIENT).

    Patient problems: No Code Available

    MAUDE report 8447877 ↗
  • InjuryMar 25, 2019Loss of Osseointegration; Loss of Osseointegration

    SMOKER, PERIODONTAL DISEASE, THREAD TAP USED, PRIMARY STABILITY WAS ACHIEVED, IMPLANT ACHIEVED OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED BY BONE, BRUXISM, BONE RESORPTION, PERI-IMPLANTITIS, SWELLING, FISTULA, ABSCESS ((B)(6) ARE SAME PATIENT).

    Patient problems: No Code Available

    MAUDE report 8447874 ↗
  • InjuryMar 20, 2019Failure to Osseointegrate; Failure to Osseointegrate

    NO OSSEOINTEGRATION.

    Patient problems: No Code Available

    MAUDE report 8435884 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page