Skip to content

SPI ELEMENT INICELL 4.23.105Recalls & reported problems

Thommen Medical AG · Class 2 · Implant

All models of SPI ELEMENT INICELL (67)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
17
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 07640156478556), received between Jul 1, 2019 and Jul 10, 2026 — classified as 17 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate13
  • Loss of Osseointegration4

Patient problems

  • Inadequate Osseointegration11
  • No Code Available6

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 10, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, INFECTION, SWELLING

    Patient problems: Inadequate Osseointegration

    MAUDE report 25775624 ↗
  • InjuryJan 13, 2026Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, SMOKER, BONE RESORPTION, PERI-IMPLANTITIS, INFECTION, PAIN, SWELLING, ABSCESS, BLEEDING.

    Patient problems: Inadequate Osseointegration

    MAUDE report 24050930 ↗
  • InjuryNov 24, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, SMOKER, INFECTION, PAIN, INCREASED SENSITIVITY, BLEEDING, INFLAMMATION, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23631064 ↗
  • InjuryAug 25, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22881897 ↗
  • InjuryMar 28, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, THREAD TAP USED, LOCAL INFECTION / SUBACUTE CHRONIC OSTEITIS, PERI-IMPLANTITIS, PAIN.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21719698 ↗
  • InjuryMar 6, 2025Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION

    Patient problems: Inadequate Osseointegration

    MAUDE report 21533566 ↗
  • InjuryJun 18, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, SMOKER, PERIODONTAL DISEASE, BRUXISM.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19556580 ↗
  • InjuryFeb 17, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION, ABSCESS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18728194 ↗
  • InjuryFeb 8, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, XEROSTOMIA, PERIODONTAL DISEASE, BRUXISM, PERI-IMPLANTITIS, INFECTION, INFLAMMATION, MOBILITY. IMPLANT 11 WAS PLACED INTO A HIGHLY COMPRIMIZED SITE. LIMITED FACIAL CORTICAL PLACE / THIN COVERING IMPLANT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18665968 ↗
  • InjuryJan 11, 2022Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERI-IMPLANTITIS, MOBILITY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 13222401 ↗
  • InjuryAug 24, 2021Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, NO THREAD TAP USED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 12358895 ↗
  • InjuryDec 9, 2020Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED.

    Patient problems: No Code Available

    MAUDE report 10973889 ↗
  • InjuryJul 27, 2020Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, SMOKER, COMPLICATION IN SITE PREPARATION.

    Patient problems: No Code Available

    MAUDE report 10324913 ↗
  • InjuryFeb 28, 2020Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERHAPS BIOMECHANICAL OVERLOAD, PAIN, MOBILITY, IMMEDIATE LOAD.

    Patient problems: No Code Available

    MAUDE report 9768179 ↗
  • InjuryJan 2, 2020Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, INFECTION, MOBILITY.

    Patient problems: No Code Available

    MAUDE report 9537309 ↗
  • InjuryJul 12, 2019Failure to Osseointegrate; Failure to Osseointegrate

    NO OSSEOINTEGRATION, THREAD TAP USED, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED BY BONE, MOBILITY. DISCOMFORT, NOT PAIN. POSSIBLY TOO HIGH OF TORQUE ACHIEVED AT IMPLANT PLACEMENT. WHEN TRYING TO UNSCREW THE HEALING CAP, THE ENTIRE IMPLANT CAME OUT WITH NO FORCE EXERTED ON THE IMPLANT.

    Patient problems: No Code Available

    MAUDE report 8784270 ↗
  • InjuryJul 1, 2019Failure to Osseointegrate; Failure to Osseointegrate

    NO THREAD TAP USED, IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WASN'T COMPLETELY COVERED BY BONE, INADEQUATE BONE QUALITY/QUANTITY.

    Patient problems: No Code Available

    MAUDE report 8747280 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates