Z5m(t) Z5m(t)-40-08Recalls & reported problems

Z-Systems AG · Monotype root-form dental implant made of Zirconia Familiy Z5m(t) Size Z5m(t)-40-08 · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Z-Systems Z5m(t)-40-08 is a zirconia root-form dental implant placed in the jaw bone. The 35 reports mostly describe the implant failing to bond with the surrounding bone, sometimes after heavy chewing forces, along with a couple of fractures and damage to surrounding tissue. Reports mention 32 injuries and no deaths. No recalls name this device, and reports have dropped from 4 in the earlier 12 months to 1 in the most recent.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
0
Injuries reported
32
Recalls naming it
0

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 07640166760504), received between Aug 31, 2020 and Feb 19, 2026 — classified as 32 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Failure to Osseointegrate” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate19
  • Loss of Osseointegration15
  • Fracture2

Patient problems

  • Unspecified Tissue Injury28
  • Tissue Damage4
  • No Clinical Signs, Symptoms or Conditions2
  • Failure of Implant1

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2026Fracture

    FRACTURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24387397 ↗
  • MalfunctionAug 22, 2025Fracture; Fracture

    FRACTURE: BIOMECHANICAL OVERLOAD.

    Patient problems: Failure of Implant

    MAUDE report 22872979 ↗
  • MalfunctionAug 22, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRAION DUE TO BIOMECHANICAL OVERLOAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22872922 ↗
  • InjuryNov 27, 2024Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 20795256 ↗
  • InjuryNov 27, 2024Failure to Osseointegrate; Failure to Osseointegrate

    NON-OSSEOINTEGRATION.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 20795027 ↗
  • InjuryAug 26, 2024Loss of Osseointegration; Loss of Osseointegration

    NON-OSSEOINTEGRATION.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 20065128 ↗
  • InjuryJul 10, 2024Loss of Osseointegration; Loss of Osseointegration

    BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 19709237 ↗
  • InjuryAug 24, 2023Loss of Osseointegration; Loss of Osseointegration

    PERI-IMPLANTITIS.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17610831 ↗
  • InjuryJul 18, 2023Failure to Osseointegrate; Failure to Osseointegrate

    INSUFFICIENT BONE QUALITY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17342320 ↗
  • InjuryJul 17, 2023Failure to Osseointegrate; Failure to Osseointegrate

    INSUFFICIENT OSSEOINTEGRATION.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17331494 ↗
  • InjuryFeb 3, 2023Loss of Osseointegration; Loss of Osseointegration

    PERI-IMPLANTITIS.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288884 ↗
  • InjuryFeb 3, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PERI-IMPLANTITIS.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288864 ↗
  • InjuryFeb 3, 2023Failure to Osseointegrate; Failure to Osseointegrate

    INSUFFICIENT BONE QUALITY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288859 ↗
  • InjuryFeb 3, 2023Loss of Osseointegration; Loss of Osseointegration

    PERI-IMPLANTITIS.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288848 ↗
  • InjuryFeb 3, 2023Failure to Osseointegrate; Failure to Osseointegrate

    INFECTION.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288847 ↗
  • InjuryFeb 3, 2023Failure to Osseointegrate; Failure to Osseointegrate

    BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288844 ↗
  • InjuryFeb 3, 2023Failure to Osseointegrate; Failure to Osseointegrate

    INSUFFICIENT BONE QUALITY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288842 ↗
  • InjuryFeb 3, 2023Failure to Osseointegrate; Failure to Osseointegrate

    INSUFFICIENT BONE QUALITY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288836 ↗
  • InjuryFeb 3, 2023Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288835 ↗
  • InjuryFeb 3, 2023Failure to Osseointegrate; Failure to Osseointegrate

    BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288771 ↗
  • InjuryDec 19, 2022Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16003039 ↗
  • InjuryDec 19, 2022Failure to Osseointegrate; Failure to Osseointegrate

    INSUFFICIENT BONE QUALITY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16003032 ↗
  • InjuryOct 14, 2022Loss of Osseointegration; Loss of Osseointegration

    PERI-IMPLANTITIS.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 15602983 ↗
  • InjuryAug 25, 2022Failure to Osseointegrate; Failure to Osseointegrate

    BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 15291549 ↗
  • InjuryMar 9, 2022Failure to Osseointegrate; Failure to Osseointegrate

    INSUFFICIENT BONE QUALITY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 13706847 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →