Skip to content

Z5m(t) Z5m(t)-50-08Recalls & reported problems

Z-Systems AG · Monotype root-form dental implant made of Zirconia Familiy Z5m(t) Size Z5m(t)-50-08 · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Z-Systems Z5m(t) zirconia dental implant is a root-form implant placed in the jawbone. Most of the 32 reports describe the implant failing to fuse with the surrounding bone, along with tissue damage around the implant site and inflammation of the gum and bone. There have been no deaths, 31 injuries and 1 malfunction, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
0
Injuries reported
31
Recalls naming it
0

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 07640166760535), received between Aug 31, 2020 and Jan 22, 2026 — classified as 31 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loss of Osseointegration” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of Osseointegration17
  • Failure to Osseointegrate15

Patient problems

  • Unspecified Tissue Injury22
  • Tissue Damage9
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 22, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24148781 ↗
  • InjuryOct 24, 2024Loss of Osseointegration; Loss of Osseointegration

    BIOMECHANICAL OVERLOAD: AFTER DEF. PROSTHETIC RESTORATION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 20520113 ↗
  • InjurySep 22, 2023Loss of Osseointegration; Loss of Osseointegration

    PERI-IMPLANTITIS.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17796942 ↗
  • InjurySep 22, 2023Loss of Osseointegration; Loss of Osseointegration

    PERI-IMPLANTITIS.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17796936 ↗
  • InjuryAug 24, 2023Loss of Osseointegration; Loss of Osseointegration

    PERI-IMPLANTITIS.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17610829 ↗
  • InjuryAug 24, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PERI-IMPLANTITIS.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17610780 ↗
  • InjuryJul 18, 2023Loss of Osseointegration; Loss of Osseointegration

    BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17342373 ↗
  • InjuryJul 18, 2023Failure to Osseointegrate; Failure to Osseointegrate

    INSUFFICIENT BONE QUALITY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17342345 ↗
  • InjuryFeb 3, 2023Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288852 ↗
  • InjuryFeb 3, 2023Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16288762 ↗
  • InjuryDec 19, 2022Failure to Osseointegrate; Failure to Osseointegrate

    INSUFFICIENT BONE QUALITY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 16003024 ↗
  • InjuryNov 29, 2022Loss of Osseointegration; Loss of Osseointegration

    INSUFFICIENT BONE QUALITY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 15876469 ↗
  • InjuryMar 9, 2022Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 13707913 ↗
  • InjuryJan 13, 2022Loss of Osseointegration; Loss of Osseointegration

    BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 13247427 ↗
  • InjuryJul 22, 2021Loss of Osseointegration; Loss of Osseointegration

    IMPLANT DID NOT INEGRATE TO BONE DUE TO BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 12210113 ↗
  • InjuryMar 29, 2021Failure to Osseointegrate; Failure to Osseointegrate

    BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11580968 ↗
  • InjuryMar 19, 2021Failure to Osseointegrate; Failure to Osseointegrate

    BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11522993 ↗
  • InjuryMar 19, 2021Loss of Osseointegration; Loss of Osseointegration

    BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11522990 ↗
  • InjuryMar 19, 2021Loss of Osseointegration; Loss of Osseointegration

    BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11522989 ↗
  • InjuryMar 19, 2021Loss of Osseointegration; Loss of Osseointegration

    INSUFFICIENT BONE QUALITY.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11522974 ↗
  • InjuryMar 19, 2021Failure to Osseointegrate; Failure to Osseointegrate

    BIOMECHANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11522942 ↗
  • InjuryFeb 8, 2021Failure to Osseointegrate; Failure to Osseointegrate

    BIOMECANICAL OVERLOAD.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11291866 ↗
  • InjuryFeb 8, 2021Failure to Osseointegrate; Failure to Osseointegrate

    PROBLEMS OF OSSEINTEGRATION CAUSED OF TISSUE AROUND IMPLANT.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11291859 ↗
  • InjuryOct 20, 2020Failure to Osseointegrate; Failure to Osseointegrate

    BIOMECHANICAL OVERLOAD.

    Patient problems: Tissue Damage

    MAUDE report 10703189 ↗
  • InjuryOct 20, 2020Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT DID NOT INTEGRATE.

    Patient problems: Tissue Damage

    MAUDE report 10703181 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates