Skip to content

Z5-BL Z5-BL-3608Recalls & reported problems

Z-Systems AG · Implant + TP-SNS + BL-HC · Class 2 · Implant

All models of Z5-BL (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07640166768463), received between Aug 13, 2024 and Sep 30, 2026 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Failure to Osseointegrate” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate6
  • Fracture1
  • Loss of Osseointegration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Implant Pain2
  • Unspecified Tissue Injury1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Fracture

    Fracture: biomechanical overload.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943207 ↗
  • MalfunctionMay 18, 2026Failure to Osseointegrate

    NON-OSSEOINTEGRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25203122 ↗
  • MalfunctionMar 26, 2026Failure to Osseointegrate

    NON-OSSEOINTEGRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24706557 ↗
  • MalfunctionJun 4, 2025Failure to Osseointegrate; Failure to Osseointegrate

    NON-OSSEOINTEGRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22139416 ↗
  • MalfunctionApr 30, 2025Failure to Osseointegrate; Failure to Osseointegrate

    NON-OSSEOINTEGRATION DUE TO INSUFFICIENT/ POOR BONE QUALITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21932514 ↗
  • MalfunctionApr 17, 2025Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION DUE TO BRUXISM.

    Patient problems: Implant Pain

    MAUDE report 21857372 ↗
  • MalfunctionMar 18, 2025Failure to Osseointegrate; Failure to Osseointegrate

    NON-OSSEOINTEGRATION DUE TO BONE AUGMENTATION.

    Patient problems: Implant Pain

    MAUDE report 21631766 ↗
  • InjuryAug 13, 2024Failure to Osseointegrate; Failure to Osseointegrate

    FRACTURE DUE TO INCORRECT POSITION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 19971084 ↗

Other Implant, Endosseous, Root-Form devices