Skip to content

Lokomat Pro 6.3 FD ADRecalls & reported problems

Hocoma AG · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07640174660414), received between Aug 6, 2026 and Aug 6, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1

Patient problems

  • Bone Fracture(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 6, 2026Improper or Incorrect Procedure or Method

    During the sixth session with the Lokomat, the patient expressed a desire to increase her walking duration to 40 minutes. At approximately 35 minutes into the session, she began to exhibit fatigable clonus in the right calf muscle complex. Despite this, she remained able to achieve adequate foot clearance with the assistance of foot lifters. The patient reported that her foot felt "weird"; at that point, clonus was explained to her, and reassurance was provided that this can occur in the context of fatigue. As the patient was close to achieving her session goal and she was encouraged to continue walking safely with support, the decision was made to continue for a further 2-3 minutes before resting. At approximately 38 minutes, the Lokomat stumble detection / cut-off feature activated, leaving the patient's right foot trapped behind her, positioned in plantar flexion and inversion within the foot lifter.

    Patient problems: Bone Fracture(s)

    MAUDE report 50156509 ↗

Other Exerciser, Powered devices