SPI®NEVO RC INICELL®, healing cap incl., PF 4.5, endosseous Ø 4.2 mm, L 9.5 mm 4.23.033Recalls & reported problems
Thommen Medical AG · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This is a Thomomen Medical SPINEVO RC Inicell dental implant, a screw-shaped root-form implant used to anchor a replacement tooth in the jaw. The 38 reports mostly describe the implant failing to fuse with the surrounding bone, and sometimes becoming loose, with patients mentioning pain, inflammation, bleeding and gum infection around the implant site. All 38 reports involve injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 39
- Deaths reported
- 0
- Injuries reported
- 39
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 07640182643485), received between Jun 21, 2024 and Sep 15, 2026 — classified as 39 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (34 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate34
- Loss of Osseointegration4
- Loosening of Implant Not Related to Bone-Ingrowth1
Patient problems
- Inadequate Osseointegration38
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.
InjurySep 15, 2026Failure to Osseointegrate
Primary stability not achieved, implant was completely covered with bone, diabetes mellitus, immediate implantation, inadequate bone quality/quantity (b)(6) are same patient).
Patient problems: Inadequate Osseointegration
MAUDE report 50770814 ↗InjuryJul 14, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION ((B)(6) AND (B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 25811949 ↗InjuryMay 19, 2026Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PERI-IMPLANTITIS, PAIN, INFLAMMATION.
Patient problems: Inadequate Osseointegration
MAUDE report 25214797 ↗InjuryMay 19, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PAIN, MOBILITY - IMPLANT WASN'T RETURNED, ONLY ABUTMENT.
Patient problems: Inadequate Osseointegration
MAUDE report 25214792 ↗InjuryMay 19, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PAIN, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 25214789 ↗InjuryMay 19, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, DIABETES MELLITUS, BLEEDING.
Patient problems: Inadequate Osseointegration
MAUDE report 25214778 ↗InjuryApr 29, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, SMOKER, PERI-IMPLANTITIS, PAIN, NUMBNESS, INFLAMMATION, MOBILITY ((B)(6) AND (B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 25026709 ↗InjuryApr 28, 2026Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED - ANOTHER SIZE OF IMPLANT WAS CHOSEN.
Patient problems: Inadequate Osseointegration
MAUDE report 25015694 ↗InjuryApr 28, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION.
Patient problems: Inadequate Osseointegration
MAUDE report 25014317 ↗InjuryApr 10, 2026Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, IMMEDIATE IMPLANTATION, PAIN.
Patient problems: Inadequate Osseointegration
MAUDE report 24838812 ↗InjuryApr 8, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED.
Patient problems: Inadequate Osseointegration
MAUDE report 24813439 ↗InjuryMar 26, 2026Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, INADEQUATE BONE QUALITY/QUANTITY, MOBILITY (B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 24697489 ↗InjuryMar 16, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY - FIBROINTEGRATION.
Patient problems: Inadequate Osseointegration
MAUDE report 24604290 ↗InjuryFeb 27, 2026Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PAIN.
Patient problems: Inadequate Osseointegration
MAUDE report 24467490 ↗InjuryJan 26, 2026Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 24176589 ↗InjuryDec 30, 2025Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PAIN ((B)(6) ARE SAME PATIENT).
Patient problems: Inadequate Osseointegration
MAUDE report 23921240 ↗InjuryDec 30, 2025Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED - ANOTHER SIZE OF IMPLANT WAS CHOSEN.
Patient problems: Inadequate Osseointegration
MAUDE report 23921201 ↗InjuryDec 29, 2025Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE.
Patient problems: Inadequate Osseointegration
MAUDE report 23905119 ↗InjuryNov 28, 2025Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 23670604 ↗InjuryNov 28, 2025Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 23670033 ↗InjuryNov 19, 2025Loss of Osseointegration; Loss of Osseointegration
LOSS OF OSSEOINTEGRATION, IMPLANT ACHIEVED OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, FISTULA.
Patient problems: Inadequate Osseointegration
MAUDE report 23591279 ↗InjuryNov 19, 2025Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, TRAUMA / ACCIDENT, BIOMECHANICAL OVERLOAD, INCREASED SENSITIVITY, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 23590915 ↗InjuryNov 5, 2025Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, PRIMARY STABILITY WAS ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, IMMEDIATE IMPLANTATION, PRECEDING/SIMULTANEOUS BONE AUGMENTATION, PAIN, INCREASED SENSITIVITY, BLEEDING, INFLAMMATION, MOBILITY.
Patient problems: Inadequate Osseointegration
MAUDE report 23475493 ↗InjuryNov 5, 2025Failure to Osseointegrate; Failure to Osseointegrate
PRIMARY STABILITY NOT ACHIEVED, IMPLANT WASN'T COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, IMMEDIATE IMPLANTATION.
Patient problems: Inadequate Osseointegration
MAUDE report 23475442 ↗InjurySep 17, 2025Failure to Osseointegrate; Failure to Osseointegrate
IMPLANT DIDN'T ACHIEVE OSSEOINTEGRATION, IMPLANT WAS COMPLETELY COVERED WITH BONE, NO THREAD TAP USED, PRECEDING/SIMULTANEOUS BONE AUGMENTATION, PAIN.
Patient problems: Inadequate Osseointegration
MAUDE report 23072785 ↗
Other Implant, Endosseous, Root-Form devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.