Skip to content

PERLA ® TL TLF-IN 05 41-NRecalls & reported problems

SpineArt SA · FINAL TIGHTENER · Class 1

All models of PERLA TL (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07640450868954), received between Mar 27, 2024 and Nov 1, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem2
  • Device Contamination with Chemical or Other Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 1, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    ON 10.OCT.2024, WE RECEIVED A COMPLAINT FROM FRANCE REGISTERED UNDER CPT-3260, REPORTING THAT DURING A SURGERY ON (B)(6) 2024, WHEN USING THE FINAL TIGHTENER INSTRUMENT REFERENCE TLF-IN 05 41-N, BATCH NUMBER 6-8534/001, WATER LEAKED INTO THE PATIENT. THE SURGEON COMPLETED THE SURGERY BY USING ANOTHER INSTRUMENT. THE SURGICAL SITE WAS WASHED WITH BETADINE SERUM (THE PATIENT IS UNDER SYSTEMATIC PREOPERATIVE ANTIBIOTIC TREATMENT). THE PATIENT IS FINE AND HE LEFT THE HOSPITAL WITH NO PROLONGED HOSPITALIZATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20586640 ↗
  • MalfunctionMar 27, 2024Contamination /Decontamination Problem; Device Contamination with Chemical or Other Material; Contamination /Decontamination Problem; Device Contamination with Chemical or Other Material

    ON (B)(6) 2024, WE RECEIVED A COMPLAINT FROM SPAIN, REGISTERED UNDER CPT-2883, REPORTING THAT DURING A SURGERY ON (B)(6) 2024, WHEN USING THE FINAL TIGHTENER INSTRUMENT REFERENCE TLF-IN 05 41-N, BATCH NUMBER 6-9649/051, BROWN WATER LEAKED INTO THE SURGICAL SITE.HIGH RISK OF INFECTION REPORTED.THE EVENT IS REPORTED TO FDA BECAUSE THIS INSTRUMENT IS USED IN THE US MARKET. THE PATIENT RECEIVED ANTIBIOTIC TREATMENT TO PREVENT ANY POTENTIAL INFECTION. NO FURTHER COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Infection

    MAUDE report 18986110 ↗

Other Screwdriver devices

All Screwdriver devices →