Skip to content

Integron® PCD26-CRecalls & reported problems

Terragene S.A. · 25 Test pack for Steam sterilization processes. · Class 2

All models of Integron (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Dec 7, 2023Class IIOpen, Classified

    Due to a discrepancy between the FDA-Cleared Indications for Use (IFU) and the actual IFU distributed with the products.

    Z-0836-2024 · Root cause: Labeling False and Misleading · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code JOJ (indicator, physical/chemical sterilization process) under 21 CFR 880.2800 as a Class II device. 21 CFR 880.2800 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Integron® PCD26-C

Has the Integron® PCD26-C been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0836-2024 (Class II), began on Dec 7, 2023: Due to a discrepancy between the FDA-Cleared Indications for Use (IFU) and the actual IFU distributed with the products. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Who makes the Integron® PCD26-C, and is it still sold?

The FDA's device database lists Terragene S.A. as the company responsible for it (brand Integron®, model PCD26-C). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Indicator, Physical/Chemical Sterilization Process devices

All Indicator, Physical/Chemical Sterilization Process devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page