Neodent 109.325Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · Ti ALVIM IMPLANT · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 07898237567105), received between Nov 9, 2017 and Dec 5, 2025 — classified as 20 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate15
  • Loss of Osseointegration3
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1

Patient problems

  • Tissue Damage12
  • Pain9
  • Unspecified Infection2
  • Complaint, Ill-Defined1
  • Fistula1
  • Swelling1
  • Swelling/ Edema1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 5, 2025Fracture; Fracture

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED IN FDI 16. ON (B)(6) 2025, FRACTURE OF THE IMPLANT WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fistula

    MAUDE report 23732278 ↗
  • InjuryJan 27, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN THE PATIENT'S MOUTH. ON (B)(6) 2021, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 13368673 ↗
  • InjuryJan 13, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN THE PATIENT'S MOUTH. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 13254086 ↗
  • InjuryJan 13, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN THE PATIENT'S MOUTH. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 13254072 ↗
  • InjuryJan 13, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN THE PATIENT'S MOUTH. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 13254070 ↗
  • InjurySep 9, 2020Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2020 IN FDI 36. ON (B)(6) 2020, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Swelling; Swelling/ Edema

    MAUDE report 10509593 ↗
  • InjuryAug 20, 2020Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 11. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2020, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 10432637 ↗
  • InjuryFeb 17, 2020Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 1 MONTH AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 30N.CM OF PRIMARY STABILITY WAS ACHIEVED, THE PATIENT PRESENTED BONE TYPE II AND IMMEDIATE IMPLANT WAS PERFORMED.

    Patient problems: Tissue Damage

    MAUDE report 9716689 ↗
  • InjuryFeb 11, 2020Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 21 DAYS THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED.

    Patient problems: Tissue Damage

    MAUDE report 9695551 ↗
  • InjuryOct 28, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 3 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED.

    Patient problems: Tissue Damage

    MAUDE report 9247205 ↗
  • InjuryApr 8, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT ALMOST 4 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED.

    Patient problems: Tissue Damage

    MAUDE report 8490867 ↗
  • InjuryJan 4, 2019Loss of Osseointegration; Loss of Osseointegration

    (B)(4) ¿ THE DENTIST REPORTED THAT 6 YEARS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS LOSS OF OSSEOINTEGRATION WAS OBSERVED. MOREOVER, THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Tissue Damage; Unspecified Infection

    MAUDE report 8218829 ↗
  • InjuryOct 1, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    (B)(4) ¿ THE DENTIST REPORTED THAT 9 MONTHS AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 15#, IT HAD TO BE REMOVED BECAUSE OF SINUS INFECTION. IN ADDITION, 50 NCM OF PRIMARY STABILITY WAS ACHIEVED, THE PATIENT PRESENTED BONE TYPE IV AND BONE GRAFT WAS EXECUTED.

    Patient problems: Pain; Tissue Damage

    MAUDE report 7923457 ↗
  • InjuryAug 29, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT AFTER DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Tissue Damage

    MAUDE report 7829901 ↗
  • InjuryNov 15, 2017Loss of Osseointegration; Loss of Osseointegration

    (B)(4). THE DENTIST REPORTED THAT, 6 YEARS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #19, ITS LOSS OF OSSEOINTEGRATION WAS OBSERVED. 60 NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Pain; Tissue Damage

    MAUDE report 7035325 ↗
  • InjuryNov 14, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT, 7 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #11; ITS NON-OSSEOINTEGRATION WAS OBSERVED. FORTY-THREE NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY (BONE TYPE II) AND BONE GRAFT WAS EXECUTED.

    Patient problems: Pain; Tissue Damage

    MAUDE report 7030862 ↗
  • InjuryNov 14, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THE LOSS OF THE DENTAL IMPLANT, BUT DID NOT PROVIDE INFORMATION ABOUT THE CASE.

    Patient problems: Complaint, Ill-Defined

    MAUDE report 7030470 ↗
  • InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT, AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Tissue Damage

    MAUDE report 7022235 ↗
  • InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT, AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Tissue Damage

    MAUDE report 7021033 ↗
  • InjuryNov 9, 2017Failure to Osseointegrate; Failure to Osseointegrate

    OS 111523 - THE DENTIST REPORTED THE NON-OSSEOINTEGRATION OF A DENTAL IMPLANT INSTALLED IN INTRAORAL REGION #3, AFTER PROSTHESIS INSTALLATION. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Tissue Damage

    MAUDE report 7012592 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →