This Neodent zygomatic dental implant is a root-form implant used to anchor replacement teeth in the upper jawbone. The 28 reports describe implants that failed to fuse with the bone or later lost that fusion, often leaving patients with pain, bleeding, tissue damage, swelling around the implant, and infection; no deaths were reported. None of these reports fall in the past 12 months, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
28
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 07898237567792), received between Nov 9, 2017 and Mar 1, 2024 — classified as 28 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (16 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate16
  • Loss of Osseointegration12

Patient problems

  • Pain18
  • Tissue Damage14
  • Unspecified Infection9
  • Fibrosis6
  • Hemorrhage/Blood Loss/Bleeding5
  • No Clinical Signs, Symptoms or Conditions4
  • Fistula3
  • Failure of Implant2

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 1, 2024Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2022 IN FDI 14. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 18819847 ↗
  • InjuryDec 28, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 16. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18413057 ↗
  • InjuryJan 30, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 26. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 16267015 ↗
  • InjuryJan 30, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 23. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 16267003 ↗
  • InjuryJul 25, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN FDI 15. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 15097391 ↗
  • InjuryJul 25, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN FDI 26. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 15097359 ↗
  • InjuryJul 25, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN FDI 13. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 15097346 ↗
  • InjuryMar 20, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI OTHERS. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13829001 ↗
  • InjuryMar 2, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN THE PATIENT'S MOUTH. ON (B)(6) 2020, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 13654481 ↗
  • InjuryMar 2, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN THE PATIENT'S MOUTH. ON (B)(6) 2020, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 13654475 ↗
  • InjuryMar 2, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN THE PATIENT'S MOUTH. ON (B)(6) 2020, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 13654450 ↗
  • InjuryFeb 28, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN THE PATIENT'S MOUTH. ON (B)(6) 2020, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 13623129 ↗
  • InjuryFeb 28, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN THE PATIENT'S MOUTH. ON (B)(6) 2020, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 13623097 ↗
  • InjuryFeb 25, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN THE PATIENT'S MOUTH. ON (B)(6) 2020, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 13604242 ↗
  • InjuryAug 31, 2020Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN THE PATIENT'S MOUTH. ON (B)(6) 2020, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Tissue Damage

    MAUDE report 10473302 ↗
  • InjuryAug 31, 2020Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN THE PATIENT'S MOUTH. ON (B)(6) 2020, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Tissue Damage

    MAUDE report 10473249 ↗
  • InjuryJan 6, 2020Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 6 MONTHS AND 10 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, IMMEDIATE LOAD PROCEDURE WAS DONE.

    Patient problems: Tissue Damage

    MAUDE report 9553743 ↗
  • InjuryApr 12, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 5 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. MOREOVER, THE PATIENT PRESENTED BONE TYPE II, BONE GRAFT WAS PLACED AND FENESTRATION HAS OCCURRED.

    Patient problems: Fistula; Pain; Tissue Damage

    MAUDE report 8511528 ↗
  • InjuryApr 12, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 5 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. MOREOVER, THE PATIENT PRESENTED BONE TYPE II, BONE GRAFT WAS PLACED AND FENESTRATION HAS OCCURRED.

    Patient problems: Fistula; Pain; Tissue Damage

    MAUDE report 8511524 ↗
  • InjuryOct 16, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT 10 MONTHS AFTER THE DENTAL IMPLANT AND PROVISIONAL ABUTMENT WERE PLACED IN ADA SITE 3 IN THE MOUTH, THE IMPLANT FAILED TO ACHIEVE OSSEOINTEGRATION. CLINICIAN NOTED PATIENT'S PERI- IMPLANTITIS, PAIN, MOBILITY AND BRUXISM. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED POST- OPERATIVE COMPLICATIONS.

    Patient problems: Pain; Tissue Damage; Unspecified Infection

    MAUDE report 7971386 ↗
  • InjuryOct 12, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT 9 MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT FOR A FIXED TEMP BRIDGE WERE PLACED IN THE RIGHT MESIAL AREA IN THE MOUTH, THE IMPLANT FAILED TO INTEGRATE IN TYPE IV BONE. CLINICIAN NOTED PATIENT'S MOBILITY. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: Failure of Implant; Tissue Damage

    MAUDE report 7961730 ↗
  • InjuryOct 12, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT 9 MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT FOR A FIXED TEMP BRIDGE WERE PLACED IN THE LEFT MESIAL AREA IN THE MOUTH, THE IMPLANT FAILED TO INTEGRATE IN TYPE IV BONE. CLINICIAN NOTED PATIENT'S MOBILITY. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: Failure of Implant; Tissue Damage

    MAUDE report 7961726 ↗
  • InjuryFeb 8, 2018Failure to Osseointegrate; Failure to Osseointegrate

    " (B)(4) - THE DENTIST REPORTED THAT 20 DAYS AFTER DENTAL IMPLANT WAS INSTALLED IN MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. A 35 NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITH IMMEDIATELY LOAD. THE PATIENT PRESENTED BONE TYPE IV. "

    Patient problems: Sinus Perforation; Tissue Damage; Unspecified Infection

    MAUDE report 7251533 ↗
  • InjuryNov 16, 2017Loss of Osseointegration; Loss of Osseointegration

    (B)(4). THE DENTIST REPORTED THAT, 9 MONTHS AFTER THE DENTAL IMPLAN WAS INSTALLED IN PATIENT¿S MOUTH; ITS NON-OSSEOINTEGRATION WAS OBSERVED. IMMEDIATELY LOAD WAS PERFORMED AND THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Pain; Tissue Damage

    MAUDE report 7036578 ↗
  • InjuryNov 15, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT, 1 YEAR AND 2 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON- OSSEOINTEGRATION WAS OBSERVED. A 50 NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS IMMEDIATELY LOADED. THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain; Tissue Damage

    MAUDE report 7035250 ↗