Neodent 109.665Recalls & reported problems
JJGC Indústria e Comércio de Materiais Dentários S/A. · CM TITAMAX EX IMPLANT · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 07898237569123), received between Nov 9, 2017 and Feb 18, 2025 — classified as 14 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate12
- Adverse Event Without Identified Device or Use Problem1
- Loss of Osseointegration1
Patient problems
- Tissue Damage9
- Pain8
- Unspecified Infection4
- Fistula3
- Fibrosis1
- No Clinical Signs, Symptoms or Conditions1
- Swelling1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryFeb 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN FDI 11, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21403752 ↗InjuryJun 23, 2023Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN THE PATIENT'S MOUTH. ON (B)(6)2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Unspecified Infection
MAUDE report 17194760 ↗InjuryNov 18, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2022 IN FDI 22. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain
MAUDE report 15827985 ↗InjuryNov 18, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN FDI 15. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain
MAUDE report 15827162 ↗InjuryDec 27, 2021Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 15. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH POOR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fistula; Pain
MAUDE report 13089344 ↗InjuryNov 16, 2018Failure to Osseointegrate; Failure to Osseointegrate
(B)(4). THE DENTIST REPORTED THAT 4 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 11#, ITS NON-OSSEOINTEGRATION WAS OBSERVED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Pain; Tissue Damage
MAUDE report 8078043 ↗InjuryMar 15, 2018Failure to Osseointegrate; Failure to Osseointegrate
(B)(4) THE DENTIST REPORTED THAT 7 DAYS AFTER DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THIRTY-FIVE (35) NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE I.
Patient problems: Pain; Tissue Damage
MAUDE report 7341399 ↗InjuryNov 14, 2017Failure to Osseointegrate; Failure to Osseointegrate
(B)(4). THE DENTIST REPORTED THAT, 8 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THIRTY FIVE (35) NCM OF PRIMARY STABILITY WAS ACHIEVED AND IMMEDIATE LOAD WAS PERFORMED. THE PATIENT PRESENTED BONE TYPE II AND BONE GRAFT WAS EXECUTED.
Patient problems: Fistula; Tissue Damage; Unspecified Infection
MAUDE report 7030711 ↗InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate
(B)(4) - THE DENTIST REPORTED THAT, 2 WEEKS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THIRTY-FIVE (35) NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS IMMEDIATELY LOADED. THE PATIENT PRESENTED BONE TYPE IV.
Patient problems: Pain; Swelling; Tissue Damage
MAUDE report 7021484 ↗InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate
(B)(4) - THE DENTIST REPORTED THAT, 1 WEEK AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #31, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THIRTY-FIVE NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/ QUANTITY (BONE TYPE IV) AND BONE GRAFT WAS EXECUTED.
Patient problems: Tissue Damage; Unspecified Infection
MAUDE report 7020282 ↗InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate
(B)(4) - THE DENTIST REPORTED THAT, 1 WEEK AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #25, ITS NON-OSSEOINTEGRATION WAS OBSERVED. FORTY NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED BONE TYPE II.
Patient problems: Fistula; Tissue Damage; Unspecified Infection
MAUDE report 7020043 ↗InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate
(B)(4) - THE DENTIST REPORTED THAT 1 MONTH AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #27, IT WAS VERIFIED ITS NON- OSSEOINTEGRATION. THIRTY-TWO NCM OF PRIMARY STABILITY WAS ACHIEVED AND IMMEDIATE LOAD WAS PERFORMED. THE PATIENT PRESENTED BONE TYPE III.
Patient problems: Pain; Tissue Damage
MAUDE report 7018413 ↗InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate
(B)(4). THE DENTIST REPORTED THAT 8 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #12 IT WAS VERIFIED ITS NON- OSSEOINTEGRATION. 50 NCM OF PRIMARY STABILITY WAS ACHIEVED AND IMMEDIATE LOAD WAS PERFORMED. THE PATIENT PRESENTED BONE TYPE II.
Patient problems: Pain; Tissue Damage
MAUDE report 7018197 ↗InjuryNov 9, 2017Failure to Osseointegrate; Failure to Osseointegrate
(B)(4). THE DENTIST REPORTED THAT 7 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #27 IT WAS VERIFIED ITS NON- OSSEOINTEGRATION.35 NCM OF PRIMARY STABILITY WAS ACHIEVED AND IMMEDIATE LOAD WAS NOT PERFORMED. THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY (BONE TYPE III) AND BONE GRAFT WAS EXECUTED.
Patient problems: Tissue Damage
MAUDE report 7015803 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page