Neodent 109.687Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · CM DRIVE IMPLANT · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent CM Drive implant is a root-form dental implant used to replace missing teeth. The 45 reports mostly describe implants that fail to bond with the surrounding bone, or lose that bond years after placement, along with tissue damage, pain and infection; no deaths were reported. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
45
Deaths reported
0
Injuries reported
45
Recalls naming it
0

Summary

The FDA's MAUDE database contains 45 adverse event reports that list this exact device (DI 07898237569253), received between Nov 9, 2017 and Jan 2, 2025 — classified as 45 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (30 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate30
  • Loss of Osseointegration8
  • Adverse Event Without Identified Device or Use Problem5
  • Difficult to Remove2

Patient problems

  • Tissue Damage33
  • Pain10
  • No Clinical Signs, Symptoms or Conditions7
  • Unspecified Infection5
  • Abscess1
  • Failure of Implant1
  • Fibrosis1
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 25 most recent of 45 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 2, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2014 IN FDI 23. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21059809 ↗
  • InjuryJan 2, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2014 IN FDI 25. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21059807 ↗
  • InjuryAug 12, 2024Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2022 IN FDI 26. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19965579 ↗
  • InjuryNov 16, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2015 IN FDI 16. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 18156263 ↗
  • InjuryNov 16, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2015 IN FDI 27. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 18156261 ↗
  • InjuryJun 23, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN FDI 26. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17194674 ↗
  • InjuryOct 13, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2016 IN ADA 4. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15598710 ↗
  • InjuryOct 13, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2016 IN ADA 4. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15597840 ↗
  • InjuryMay 10, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN THE PATIENT'S MOUTH. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 11800709 ↗
  • InjuryMar 29, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN THE PATIENT'S MOUTH. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11584424 ↗
  • InjuryMar 5, 2021Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2015 IN FDI 11. ON (B)(6) 2021, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH POOR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 11425240 ↗
  • InjuryDec 15, 2020Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN FDI 36. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2019, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, MOBILITY AND ABSCESS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Abscess; Fibrosis; Unspecified Infection

    MAUDE report 11006227 ↗
  • InjuryFeb 20, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 8 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED AND OCCLUSAL OVERLOAD HAS OCCURRED. MOREOVER, 35N.CM OF PRIMARY STABILITY WAS ACHIEVED, THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY (BONE TYPE IV), BONE GRAFT WAS PLACED, IMMEDIATE IMPLANT WAS PERFORMED AND IMMEDIATE LOAD PROCEDURE WAS DONE.

    Patient problems: Tissue Damage

    MAUDE report 8355055 ↗
  • InjuryFeb 8, 2019Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED 1 YEAR AFTER THE DENTAL IMPLANT WAS PLACED IN ADA SITE 10 IN THE MOUTH, THE IMPLANT DID NOT ACHIEVE OSSEOINTEGRATION. CLINICIAN REPORTS THE PATIENT'S TRAUMA AND IMPLANT MOBILITY. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED PATIENT COMPLICATIONS. (B)(4).

    Patient problems: Tissue Damage

    MAUDE report 8324013 ↗
  • InjuryOct 22, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT ALMOST 6 MONTHS AFTER THE DENTAL IMPLANT WAS PLACED IN ADA SITE 13 IN THE MOUTH, THE IMPLANT DID NOT ACHIEVE OSSEOINTEGRATION. AFTER THIS IMPLANT WAS REMOVED, ANOTHER WAS PLACED IN THE SAME SITE. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: Failure of Implant; Tissue Damage

    MAUDE report 7993088 ↗
  • InjuryAug 29, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS 5 MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT WERE PLACED IN (B)(6) SITE 3 IN THE MOUTH, THE IMPLANT FAILED IN TYPE II BONE. CLINICIAN REPORTS PAIN. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED POST-OPERATIVE COMPLICATIONS.

    Patient problems: Pain; Tissue Damage

    MAUDE report 7830418 ↗
  • InjuryAug 29, 2018Failure to Osseointegrate; Failure to Osseointegrate

    RP.013610 ¿ THE DENTIST REPORTED THAT 7 DAYS AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 5#, ITS NON-OSSEOINTEGRATION WAS OBSERVED. 40NCM OF PRIMARY STABILITY WAS ACHIEVED, THE PATIENT PRESENTED BONE TYPE II AND IMMEDIATE IMPLANT PROCEDURE WAS PERFORMED.

    Patient problems: Tissue Damage

    MAUDE report 7829912 ↗
  • InjuryJun 6, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT 3 MONTHS AFTER THE DENTAL IMPLANT WAS PLACED IN ADA SITE 29 IN THE MOUTH, THE IMPLANT HAD NOT YET ACHIEVED OSSEOINTEGRATION IN TYPE III BONE. CLINICIAN REPORTS PAIN AND PERI- IMPLANTITIS. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED POST-OPERATIVE COMPLICATIONS.

    Patient problems: Pain; Tissue Damage; Unspecified Infection

    MAUDE report 7574675 ↗
  • InjuryApr 19, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT ALMOST 6 MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT WERE PLACED IN ADA SITE 3, THE IMPLANT HAD NOT ACHIEVED OSSEOINTEGRATION IN TYPE I BONE CLINICIAN REPORTS BONE LOSS. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED COMPLICATIONS. (B)(4).

    Patient problems: Tissue Damage

    MAUDE report 7443667 ↗
  • InjuryMar 16, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 15 DAYS AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #6, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, 60NCM OF PRIMARY STABILITY WAS ACHIEVED, THE PATIENT PRESENTED BONE TYPE III AND THE PROCEDURES OF IMMEDIATE IMPLANT AND IMMEDIATE LOAD WERE PERFORMED.

    Patient problems: Pain; Tissue Damage

    MAUDE report 7346871 ↗
  • InjuryFeb 23, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). - THE DENTIST REPORTED THAT, 7 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #3, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THIRTY FIVE NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS IMMEDIATELY LOADED. THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Tissue Damage; Unspecified Infection

    MAUDE report 7293570 ↗
  • InjuryFeb 9, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT 4.5 MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT WERE PLACED IN ADA SITE 15, THE IMPLANT LOST OSSEOINTEGRATION. CLINICIAN REPORTS BRUXISM. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED COMPLICATIONS. (B)(4).

    Patient problems: Tissue Damage

    MAUDE report 7259062 ↗
  • InjuryFeb 7, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT 4.5 MONTHS AFTER THE DENTAL IMPLANT WAS PLACED IN ADA SITE 13, THE IMPLANT HAD NOT YET OBTAINED OSEOINTEGRATION. CLINICIAN REPORTS BRUXISM. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED COMPLICATIONS.

    Patient problems: Tissue Damage

    MAUDE report 7249698 ↗
  • InjuryDec 22, 2017Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT ALMOST 6 MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT WERE PLACED IN ADA SITE 31, NON-OSSEOINTEGRATION WAS VERIFIED IN TYPE III BONE. A BONE GRAFT WAS PERFORMED WITH BIO-OSS. CLINICIAN NOTED BRUXISM. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED POST-OPERATIVE COMPLICATIONS.

    Patient problems: Tissue Damage

    MAUDE report 7142196 ↗
  • InjuryNov 15, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    (B)(4). THE DENTIST REPORTED THAT, THE DENTAL IMPLANT WAS REMOVED DURING ITS INSTALLATION SURGERY BECAUSE NO PRIMARY STABILITY WAS ACHIEVED. THERE IS NO INFORMATION ABOUT IMPLANT REPLACEMENT IN THE SAME SURGICAL ACT.

    Patient problems: Tissue Damage

    MAUDE report 7035079 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices