Skip to content

Neodent 140.621Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · CM ALVIM ACQUA IMPLANT · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent Alvim Acqua dental implant has 232 adverse event reports, including 232 injuries and no deaths. The reports mostly describe the implant failing to bond with the bone, and patients reporting tissue damage, pain, or infection. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
233
Deaths reported
0
Injuries reported
233
Recalls naming it
0

Summary

The FDA's MAUDE database contains 233 adverse event reports that list this exact device (DI 07899878016502), received between Nov 9, 2017 and Aug 27, 2026 — classified as 233 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (192 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 4 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 3 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 5 reportsNov 2024: 6 reportsDec 2024: 1 reportsJan 2025: 3 reportsFeb 2025: 7 reportsMar 2025: 7 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 8 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 4 reportsApr 2026: 2 reportsMay 2026: 10 reportsJun 2026: 4 reportsJul 2026: 2 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate192
  • Loss of Osseointegration24
  • Adverse Event Without Identified Device or Use Problem17
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions94
  • Tissue Damage84
  • Pain57
  • Unspecified Infection33
  • Fibrosis18
  • Fistula16
  • Swelling/ Edema12
  • Inflammation8

Recent adverse event reports

The 25 most recent of 233 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2015 in FDI 16. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Pain and Swelling. No further patient complications were reported.

    Patient problems: Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 50482252 ↗
  • InjuryAug 3, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2025(b)(6) in FDI 35. On 2025(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50121474 ↗
  • InjuryJul 30, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6 )2026 in FDI 21. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain, Swelling and Inflammation. No further patient complications were reported.

    Patient problems: Inflammation; Pain; Swelling/ Edema

    MAUDE report 50089301 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2025 in FDI 43. On (B)(6) 2025, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50011335 ↗
  • InjuryJun 19, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25565381 ↗
  • InjuryJun 19, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25565305 ↗
  • InjuryJun 19, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25565290 ↗
  • InjuryJun 8, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2026 IN FDI 12. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25434079 ↗
  • InjuryMay 26, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025 IN FDI 14. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25283677 ↗
  • InjuryMay 12, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25152035 ↗
  • InjuryMay 12, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25152023 ↗
  • InjuryMay 12, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6)2025 IN THE PATIENT'S MOUTH. ON (B)(6)2025, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25152019 ↗
  • InjuryMay 11, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25141907 ↗
  • InjuryMay 11, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25141902 ↗
  • InjuryMay 11, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25141744 ↗
  • InjuryMay 11, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25141740 ↗
  • InjuryMay 11, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25141659 ↗
  • InjuryMay 6, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 11. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25102629 ↗
  • InjuryApr 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 43. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24864889 ↗
  • InjuryApr 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 33. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24864838 ↗
  • InjuryMar 16, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025 IN FDI 24. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24610857 ↗
  • InjuryMar 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24593392 ↗
  • InjuryMar 6, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH POOR ORAL HYGIENE. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN AND FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fistula; Pain; Unspecified Infection

    MAUDE report 24535969 ↗
  • InjuryMar 5, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 24523411 ↗
  • InjuryFeb 20, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 12. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24404481 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page