Skip to content

Neodent 140.664Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · CM TITAMAX EX ACQUA IMPLANT · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent CM TITAMAX EX ACQUA implant is a root-form dental implant used to replace missing teeth. The 33 reports mostly describe implants that fail to bond with the surrounding bone or lose that bond later, sometimes right at placement, along with tissue damage, pain, swelling and infection. No deaths were reported, and 33 injuries were. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
33
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 07899878016762), received between Nov 9, 2017 and May 26, 2026 — classified as 33 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (21 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate21
  • Adverse Event Without Identified Device or Use Problem7
  • Loss of Osseointegration5

Patient problems

  • Tissue Damage22
  • Pain7
  • No Clinical Signs, Symptoms or Conditions6
  • Unspecified Infection5
  • Abscess2
  • Failure of Implant2
  • Fibrosis1
  • Swelling1

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 26, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 35. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25283834 ↗
  • InjuryMar 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN THE PATIENT'S MOUTH, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED, IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND IMMEDIATE EXTRACTION SITE. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21636616 ↗
  • InjuryJan 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN FDI 46, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21142778 ↗
  • InjuryJan 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN FDI 45, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21128899 ↗
  • InjuryJan 10, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH POOR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 21128827 ↗
  • InjuryApr 17, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 12. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 16758172 ↗
  • InjurySep 27, 2021Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN ADA 4. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2021, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12535748 ↗
  • InjurySep 27, 2021Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN ADA 4. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2021, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12534079 ↗
  • InjuryJun 11, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2016 IN FDI 24. ON (B)(6) 2017, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: ABSCESS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Abscess

    MAUDE report 11986049 ↗
  • InjuryJun 11, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2016 IN FDI 14. ON (B)(6) 2017, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: ABSCESS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Abscess

    MAUDE report 11985983 ↗
  • InjuryJun 10, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2016 IN FDI 14. ON (B)(6) 2017 NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 11977027 ↗
  • InjuryJul 16, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT, AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #6, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Tissue Damage

    MAUDE report 8794948 ↗
  • InjuryNov 7, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT 1.5 MONTHS AFTER THE DENTAL IMPLANT WAS PLACED IN ADA SITE 8 IN THE MOUTH, THE IMPLANT DID NOT ACHIEVE OSSEOINTEGRATION. THE SITE WAS GRAFTED. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: Failure of Implant; Tissue Damage

    MAUDE report 8048690 ↗
  • InjuryNov 7, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT 1.5 MONTHS AFTER THE DENTAL IMPLANT WAS PLACED IN ADA SITE 24 IN THE MOUTH, THE IMPLANT DID NOT ACHIEVE OSSEOINTEGRATION. THE SITE WAS GRAFTED. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: Failure of Implant; Tissue Damage

    MAUDE report 8048689 ↗
  • InjuryOct 4, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTED THAT ON THE DAY THE IMPLANT WAS PLACED IN ADA SITE 26 IN THE MOUTH, PRIMARY STABILITY WAS NOT ACHIEVED IN TYPE III BONE. THE IMPLANT WAS REMOVED AND THE SITE WAS GRAFTED. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: Tissue Damage

    MAUDE report 7933816 ↗
  • InjuryMay 25, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). - THE DENTIST REPORTED THAT 2 MONTHS AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 5#, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, THE PATIENT PRESENTED BONE TYPE II, IMMEDIATE IMPLANT AND IMMEDIATE LOAD WERE PERFORMED AND BONE GRAFT WAS EXECUTED.

    Patient problems: Pain; Tissue Damage; Unspecified Infection

    MAUDE report 7545933 ↗
  • InjuryMay 18, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT NEARLY 6 MONTHS AFTER THE DENTAL IMPLANT AND FULL LOWER RESTORATION WERE PLACED (ADA SITE 29 IN THE MOUTH), THE IMPLANT FAILED DUE TO A BIOMECHANICAL OVERLOAD. CLINICIAN REPORTS PERI- IMPLANTITIS, BONE LOSS AND INFLAMMATION. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED POST- OPERATIVE COMPLICATIONS.

    Patient problems: Tissue Damage; Unspecified Infection

    MAUDE report 7527184 ↗
  • InjuryMay 15, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTED A FAILURE OCCURRED ON THE DAY THE IMPLANT (ADA SITE 7 IN MOUTH) AND BRIDGE WERE PLACED. CLINICIAN REPORTS PAIN AND INSUFFICIENT BONE QUALITY AND QUANTITY. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED POST- OPERATIVE COMPLICATIONS.

    Patient problems: Pain; Tissue Damage

    MAUDE report 7516427 ↗
  • InjuryApr 5, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 18 DAYS AFTER DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, THE PATIENT PRESENTED BONE TYPE II, 40NCM OF PRIMARY STABILITY WAS ACHIEVED, IMMEDIATE LOAD AND IMMEDIATE IMPLANT WERE PERFORMED AND BONE GRAFT WAS EXECUTED.

    Patient problems: Pain; Tissue Damage; Unspecified Infection

    MAUDE report 7401027 ↗
  • InjuryFeb 23, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS FOUR MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT WERE PLACED, THE IMPLANT HAD NOT ACHIEVED OSSEOINTEGRATION. THE SITE WAS GRAFTED WITH THE PATIENT'S OWN BONE. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED COMPLICATIONS.

    Patient problems: Tissue Damage

    MAUDE report 7293561 ↗
  • InjuryFeb 7, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT NEARLY 3 MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT WERE PLACED IN ADA SITE 20, THE IMPLANT WAS MOBILE. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED COMPLICATIONS.

    Patient problems: Tissue Damage

    MAUDE report 7247457 ↗
  • InjuryFeb 7, 2018Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT 3 MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT WERE PLACED IN ADA SITE 26, THE IMPLANT WAS MOBILE. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED COMPLICATIONS.

    Patient problems: Tissue Damage

    MAUDE report 7247448 ↗
  • InjuryFeb 1, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 2 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 11, ITS NON-OSSEOINTEGRATION WAS OBSERVED. 32 NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED BONE TYPE IV.

    Patient problems: Tissue Damage; Unspecified Infection

    MAUDE report 7235285 ↗
  • InjuryFeb 1, 2018Loss of Osseointegration; Loss of Osseointegration

    (B)(4). - THE DENTIST REPORTED THAT 16 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THIRTY FIVE NCM OF PRIMARY STABILITY WAS ACHIEVED THE IMPLANT WAS INSTALLED WITH IMMEDIATELY LOAD.

    Patient problems: Pain; Tissue Damage

    MAUDE report 7235253 ↗
  • InjuryFeb 1, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 13 DAYS AFTER DENTAL IMPLANT WAS INSTALLED IN MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. 35 NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Tissue Damage; Unspecified Infection

    MAUDE report 7235220 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page