JJGC Indústria e Comércio de Materiais Dentários S/A. · CM DRIVE ACQUA IMPLANT · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent CM Drive Acqua implant is a dental root-form implant used to replace missing teeth. Most of the 55 reports describe the implant failing to fuse with the jawbone or later losing that fusion, with patients reporting pain, bleeding, tissue damage, infection around the implant, and a loose implant. All 55 reports are injury reports, with no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
55
Deaths reported
0
Injuries reported
55
Recalls naming it
0

Summary

The FDA's MAUDE database contains 55 adverse event reports that list this exact device (DI 07899878016878), received between Nov 10, 2017 and Jun 18, 2026 — classified as 55 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (32 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate32
  • Adverse Event Without Identified Device or Use Problem13
  • Loss of Osseointegration10

Patient problems

  • Tissue Damage42
  • Pain10
  • Unspecified Infection10
  • No Clinical Signs, Symptoms or Conditions5
  • Fibrosis3
  • Hemorrhage/Blood Loss/Bleeding3
  • Hyperresponsive to stimuli2
  • Complaint, Ill-Defined1

Recent adverse event reports

The 25 most recent of 55 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 18, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 26. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25551351 ↗
  • InjuryNov 4, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 36. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 23467758 ↗
  • InjuryJan 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN FDI 16, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: SINUS PERFORATION AND IMMEDIATE EXTRACTION SITE. PATIENT PRESENTED WITH BONE TYPE IV. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: Sinus Perforation

    MAUDE report 21151291 ↗
  • InjuryJul 1, 2024Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN ADA 10. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND RIDGE AUGMENTATION. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, MOBILITY AND INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hyperresponsive to stimuli; Unspecified Infection

    MAUDE report 19649693 ↗
  • InjuryJul 1, 2024Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN ADA 10. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND RIDGE AUGMENTATION. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, MOBILITY AND INCREASED SENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hyperresponsive to stimuli; Unspecified Infection

    MAUDE report 19646386 ↗
  • InjuryOct 17, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN THE PATIENT'S MOUTH. ON (B)(6)2022, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 15619573 ↗
  • InjuryApr 21, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN ADA 23. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14171331 ↗
  • InjuryApr 21, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN ADA 23. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14169917 ↗
  • InjuryApr 19, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2016 IN ADA 15. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14145536 ↗
  • InjuryApr 19, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2016 IN ADA 15. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14143256 ↗
  • InjurySep 21, 2020Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN THE PATIENT'S MOUTH. ON (B)(6)2020, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 10562215 ↗
  • InjuryAug 14, 2020Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN THE PATIENT'S MOUTH. ON (B)(6) 2020, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 10409363 ↗
  • InjuryApr 23, 2020Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT ALMOST 3 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED.

    Patient problems: Tissue Damage

    MAUDE report 9993174 ↗
  • InjuryApr 8, 2020Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN THE PATIENT'S MOUTH. ON (B)(6) 2019, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: Tissue Damage

    MAUDE report 9934552 ↗
  • InjuryNov 11, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 2 MONTHS AND 8 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 22#, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Fistula; Tissue Damage

    MAUDE report 9305627 ↗
  • InjuryOct 10, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT 1 MONTH AND 19 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 13#, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Pain; Tissue Damage; Unspecified Infection

    MAUDE report 9174604 ↗
  • InjurySep 18, 2019Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT 1.5 MONTHS AFTER THE DENTAL IMPLANT WAS PLACED IN ADA SITE 13 IN THE PATIENT'S MOUTH, THE IMPLANT DID NOT ACHIEVE OSSEOINTEGRATION IN TYPE II BONE. THE CLINICIAN REPORTED AN INFECTION, PAIN, IMPLANT MOBILITY AND BLEEDING IN THIS PATIENT WITH FAIR ORAL HYGIENE. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED POST-OPERATIVE COMPLICATIONS.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain; Tissue Damage; Unspecified Infection

    MAUDE report 9084325 ↗
  • InjuryMay 9, 2019Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT 2 MONTHS AFTER THE DENTAL IMPLANT WAS PLACED IN ADA SITE 3 IN THE PATIENT'S MOUTH, THE IMPLANT DID NOT ACHIEVE OSSEOINTEGRATION. THE CLINICIAN NOTED THE IMPLANT'S MOBILITY AND THE PATIENT'S PAIN AND POOR HYGIENE. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED POST-OPERATIVE COMPLICATIONS.

    Patient problems: Pain; Tissue Damage

    MAUDE report 8597447 ↗
  • InjuryApr 26, 2019Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT OVER 1 MONTH AFTER THE IMMEDIATE DENTAL IMPLANT WAS PLACED IN ADA SITE 5 IN THE PATIENT'S MOUTH, THE IMPLANT DID NOT ACHIEVE OSSEOINTEGRATION IN TYPE II BONE. THE CLINICIAN NOTED THE PATIENT'S BRUXISM. THE IMPLANT WAS TORQUED MANUALLY TO 45 NCM. THERE WERE NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: Tissue Damage

    MAUDE report 8558233 ↗
  • InjuryApr 26, 2019Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT OVER 1 MONTH AFTER THE IMMEDIATE DENTAL IMPLANT WAS PLACED IN ADA SITE 4 IN THE PATIENT'S MOUTH, THE IMPLANT DID NOT ACHIEVE OSSEOINTEGRATION IN TYPE II BONE. THE CLINICIAN NOTED THE PATIENT'S BRUXISM. THE IMPLANT WAS TORQUED MANUALLY TO 45 NCM. THERE WERE NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: Tissue Damage

    MAUDE report 8558229 ↗
  • InjuryApr 22, 2019Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTED THAT 11 MONTHS AFTER THE DENTAL IMPLANT WAS PLACED IN ADA SITE 10 IN THE PATIENT'S MOUTH, THE IMPLANT LOST OSSEOINTEGRATION IN TYPE II BONE. THE CLINICIAN NOTED THE PATIENT'S PAIN. THE SITE WAS GRAFTED. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED POST-OPERATIVE COMPLICATIONS.

    Patient problems: Pain; Tissue Damage

    MAUDE report 8537739 ↗
  • InjuryMar 29, 2019Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTED 9.5 MONTHS AFTER THE DENTAL IMPLANT AND ABUTMENT WERE PLACED IN ADA SITE 7 IN THE MOUTH, THE IMPLANT DID NOT ACHIEVE OSSEOINTEGRATION IN TYPE II BONE. THE IMPLANT WAS TORQUED TO 50 NCM. THE IMPLANT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: Tissue Damage

    MAUDE report 8467054 ↗
  • InjuryMar 6, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT ALMOST 2 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 30N.CM OF PRIMARY STABILITY WAS ACHIEVED, THE PATIENT PRESENTED BONE TYPE IV, BONE GRAFT WAS PLACED AND IMMEDIATE IMPLANT WAS PERFORMED.

    Patient problems: Tissue Damage

    MAUDE report 8397004 ↗
  • InjuryFeb 7, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 1 MONTH AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 30N.CM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY (BONE TYPE III).

    Patient problems: Tissue Damage

    MAUDE report 8314793 ↗
  • InjuryJan 22, 2019Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTED THAT ALMOST 2 MONTHS AFTER THE DENTAL IMPLANT WAS PLACED IN ADA SITE 11 IN THE MOUTH, THE IMPLANT DID NOT INTEGRATE. CLINICIAN REPORTED PATIENT'S DISEASED MUCOUS MEMBRANE, PERI-IMPLANTITIS, SINUS PERFORATION, PAIN, MOBILITY, BLEEDING AND COMPLICATIONS IN THE SITE PREP. THE SITE WAS GRAFTED. THERE WERE NO REPORTED POST-OPERATIVE COMPLICATIONS. (B)(4).

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain; Tissue Damage; Unspecified Infection

    MAUDE report 8269147 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →