Neodent 140.738Recalls & reported problems
JJGC Indústria e Comércio de Materiais Dentários S/A. · FACILITY ACQUA IMPLANT · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 16
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 07899878017042), received between Nov 10, 2017 and Jul 28, 2023 — classified as 16 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Failure to Osseointegrate6
- Loss of Osseointegration6
- Adverse Event Without Identified Device or Use Problem4
Patient problems
- Tissue Damage8
- No Clinical Signs, Symptoms or Conditions5
- Pain3
- Unspecified Infection3
- Fibrosis2
- Fistula1
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
InjuryJul 28, 2023Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2022 IN THE PATIENT'S MOUTH. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection
MAUDE report 17416956 ↗InjuryMay 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 11. ON (B)(6) 2023, THE IMPLANT WAS REMOVED UPON CLINICIAN¿S DECISION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16924117 ↗InjuryMay 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 21. ON (B)(6) 2023, THE IMPLANT WAS REMOVED UPON CLINICIAN¿S DECISION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16924076 ↗InjuryMay 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 12. ON (B)(6) 2023, THE IMPLANT WAS REMOVED UPON CLINICIAN¿S DECISION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16924068 ↗InjuryOct 31, 2022Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN FDI 44. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15702918 ↗InjuryJan 10, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 22. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION AND FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fistula; Unspecified Infection
MAUDE report 13210327 ↗InjuryJun 7, 2021Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2021 LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain; Unspecified Infection
MAUDE report 11950483 ↗InjuryJan 25, 2021Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN THE PATIENT'S MOUTH. ON (B)(6) 2020, LOSS OF OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 11225065 ↗InjurySep 14, 2020Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 32. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2020, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH POOR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions; Tissue Damage
MAUDE report 10530802 ↗InjuryApr 24, 2020Loss of Osseointegration; Loss of Osseointegration
(B)(4) ¿ THE DENTIST REPORTED THAT ALMOST 7 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 10#, ITS LOSS OF OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 35N.CM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE III.
Patient problems: Tissue Damage
MAUDE report 9996677 ↗InjuryOct 3, 2019Failure to Osseointegrate; Failure to Osseointegrate
(B)(4) - THE DENTIST REPORTED THAT 5 MONTHS AND 9 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 25#, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, THE PATIENT PRESENTED BONE TYPE III.
Patient problems: Pain; Tissue Damage
MAUDE report 9149994 ↗InjurySep 12, 2019Failure to Osseointegrate; Failure to Osseointegrate
(B)(4) - THE DENTIST REPORTED THAT 2 MONTHS AND 11 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, THE PATIENT PRESENTED BONE TYPE III.
Patient problems: Tissue Damage
MAUDE report 9008309 ↗InjuryAug 22, 2019Loss of Osseointegration; Loss of Osseointegration
(B)(4)- THE DENTIST REPORTED THAT ALMOST 7 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 9#, ITS LOSS OF OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 20N.CM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE III. IMMEDIATE LOAD PROCEDURE WAS PERFORMED AND BONE GRAFT WAS PLACED.
Patient problems: Tissue Damage
MAUDE report 8921718 ↗InjuryJul 15, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
(B)(4). THE DENTIST REPORTED THAT HE REMOVED THE DENTAL IMPLANT DURING ITS INSTALLATION SURGERY BECAUSE THE IMPLANT DRIVER FRACTURED INSIDE OF THE IMPLANT. THERE IS NO INFORMATION ABOUT IMPLANT REPLACEMENT IN THE SAME SURGICAL ACT.
Patient problems: Tissue Damage
MAUDE report 8788687 ↗InjuryDec 20, 2018Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTED THAT 2.5 MONTHS AFTER THE DENTAL IMPLANT AND HEALING CAP WERE PLACED IN ADA SITE 12 IN THE MOUTH, THE IMPLANT DID NOT ACHIEVE OSSEOINTEGRATION IN TYPE II BONE. THE PRODUCT WILL BE FORWARDED TO THE MANUFACTURER FOR INVESTIGATION. THERE WERE NO REPORTED PATIENT COMPLICATIONS.
Patient problems: Tissue Damage
MAUDE report 8185456 ↗InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate
(B)(4). THE DENTIST REPORTED THAT, 4 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #10, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD AND THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY (BONE TYPE II).
Patient problems: Tissue Damage
MAUDE report 7021423 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page