Neodent 140.752Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · Ti DRIVE ACQUA IMPLANT · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent Ti Drive Acqua implant is a root-form dental implant used to anchor replacement teeth in the jaw. The 44 reports mostly describe implants that fail to bond with the surrounding bone, or that lose that bond after months or years, sometimes with pain, swelling, or damage to surrounding tissue. No deaths were reported, and all 44 reports involved injuries. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
44
Deaths reported
0
Injuries reported
44
Recalls naming it
0

Summary

The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 07899878017073), received between Nov 8, 2017 and Jan 20, 2025 — classified as 44 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (34 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate34
  • Loss of Osseointegration10

Patient problems

  • Tissue Damage38
  • Pain10
  • Fistula2
  • No Clinical Signs, Symptoms or Conditions2
  • Swelling2
  • Unspecified Infection2
  • Fibrosis1
  • Swelling/ Edema1

Recent adverse event reports

The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 20, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN THE PATIENT'S MOUTH. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21192839 ↗
  • InjuryApr 5, 2024Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 15. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH POOR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 19053271 ↗
  • InjuryJan 11, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2017 IN THE PATIENT'S MOUTH. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16142683 ↗
  • InjuryOct 10, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 31. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 15572221 ↗
  • InjuryJan 21, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN THE PATIENT'S MOUTH. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 13317040 ↗
  • InjuryNov 25, 2021Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN FDI 32. ON (B)(6) 2021, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 12875241 ↗
  • InjuryJul 20, 2020Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN THE PATIENT'S MOUTH. ON (B)(6) 2019, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Tissue Damage

    MAUDE report 10297641 ↗
  • InjuryJul 20, 2020Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN THE PATIENT'S MOUTH. ON (B)(6) 2019, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Tissue Damage

    MAUDE report 10297619 ↗
  • InjuryAug 19, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 1 MONTH AND 5 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 20#, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 32N.CM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Tissue Damage

    MAUDE report 8905876 ↗
  • InjuryMay 24, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 5 MONTHS AND 23 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 29#, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 40N.CM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Tissue Damage

    MAUDE report 8639760 ↗
  • InjuryMar 28, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 6 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 6#, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 30N.CM OF PRIMARY STABILITY WAS ACHIEVED, THE PATIENT PRESENTED BONE TYPE III AND BONE GRAFT WAS PLACED.

    Patient problems: Tissue Damage

    MAUDE report 8462186 ↗
  • InjuryMar 11, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 4 MONTHS AND 3 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 18#, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 20N.CM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY (BONE TYPE II).

    Patient problems: Tissue Damage

    MAUDE report 8407506 ↗
  • InjuryJan 17, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 6 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Pain; Tissue Damage

    MAUDE report 8257264 ↗
  • InjuryJan 15, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 3 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Tissue Damage

    MAUDE report 8249316 ↗
  • InjuryJan 15, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT ALMOST 1 YEAR AND 2 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON- OSSEOINTEGRATION WAS OBSERVED.

    Patient problems: Tissue Damage

    MAUDE report 8249310 ↗
  • InjuryOct 4, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 29#, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, 45NCM OF PRIMARY STABILITY WAS ACHIEVED.

    Patient problems: Swelling; Tissue Damage; Unspecified Infection

    MAUDE report 7936207 ↗
  • InjuryOct 4, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 2 MONTHS AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 4#, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, 30NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Tissue Damage

    MAUDE report 7936189 ↗
  • InjuryAug 29, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 4#, ITS NON-OSSEOINTEGRATION WAS OBSERVED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Tissue Damage

    MAUDE report 7829143 ↗
  • InjuryApr 12, 2018Loss of Osseointegration; Loss of Osseointegration

    (B)(6) - THE DENTIST REPORTED THAT 3 YEARS AND 4 MONTHS AFTER DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Fistula; Pain; Tissue Damage

    MAUDE report 7425352 ↗
  • InjuryApr 12, 2018Loss of Osseointegration; Loss of Osseointegration

    (B)(4). THE DENTIST REPORTED THAT 3 YEARS AND 4 MONTHS AFTER DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Fistula; Pain; Tissue Damage

    MAUDE report 7425349 ↗
  • InjuryFeb 2, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT 2 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. FORTY NCM OF PRIMARY STABILITY WAS ACHIEVED THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Tissue Damage

    MAUDE report 7237271 ↗
  • InjuryDec 21, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT, 1 MONTH AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #5, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THIRTY NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Tissue Damage

    MAUDE report 7141176 ↗
  • InjuryNov 17, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT, 3 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #5, ITS NON-OSSEOINTEGRATION WAS OBSERVED. TWENTY NCM OF PRIMARY STABILITY WAS ACHIEVED. THE PATIENT PRESENTED INSUFFICIENT BONE QUALITYQUANTITY (BONE TYPE III), FENESTRATION OCCURRED DURING SURGERY AND BONE GRAFT WAS EXECUTED.

    Patient problems: Tissue Damage

    MAUDE report 7041523 ↗
  • InjuryNov 15, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT, 3 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #20, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THE 45 NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Pain; Swelling; Tissue Damage

    MAUDE report 7032382 ↗
  • InjuryNov 14, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT, 7 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #19, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD AND THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/ QUANTITY (BONE TYPE II).

    Patient problems: Pain; Tissue Damage

    MAUDE report 7030680 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →