Neodent 140.754Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · Ti DRIVE ACQUA IMPLANT · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
18
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 07899878017097), received between Nov 10, 2017 and Jun 17, 2025 — classified as 18 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate11
  • Loss of Osseointegration4
  • Difficult to Remove2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Tissue Damage14
  • Pain4
  • Unspecified Infection3
  • Fistula2
  • Fibrosis1
  • Inflammation1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 17, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2014 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN, FISTULA AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fistula; Inflammation; Pain

    MAUDE report 22233856 ↗
  • InjuryJan 20, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2015 IN THE PATIENT'S MOUTH. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 21192838 ↗
  • InjurySep 21, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2023 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 17793791 ↗
  • InjuryMay 23, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN FDI 12. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fistula; Pain

    MAUDE report 14473741 ↗
  • InjuryNov 7, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 2 MONTHS AND 11 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 32N.CM OF PRIMARY STABILITY WAS ACHIEVED, THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Tissue Damage

    MAUDE report 9291579 ↗
  • InjuryJul 26, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT, 3 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #10, ITS NON-OSSEOINTEGRATION WAS OBSERVED. 35 NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Tissue Damage

    MAUDE report 8832593 ↗
  • InjuryJul 26, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT, 3 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #7, ITS NON-OSSEOINTEGRATION WAS OBSERVED. 35 NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Tissue Damage

    MAUDE report 8832592 ↗
  • InjuryJul 3, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 45N.CM OF PRIMARY STABILITY WAS ACHIEVED AND IMMEDIATE IMPLANT PROCEDURE WAS PERFORMED.

    Patient problems: Tissue Damage

    MAUDE report 8757330 ↗
  • InjuryAug 29, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 3 MONTHS AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 13#, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, 35NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Tissue Damage

    MAUDE report 7829958 ↗
  • InjuryAug 29, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 5 MONTHS AFTER DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THE PATIENT PRESENTED BONE TYPE III AND 40NCM OF PRIMARY STABILITY WAS ACHIEVED.

    Patient problems: Tissue Damage

    MAUDE report 7829311 ↗
  • InjuryAug 29, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT AFTER DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Tissue Damage

    MAUDE report 7829127 ↗
  • InjuryMay 3, 2018Difficult to Remove; Difficult to Remove

    (B)(4) - THE DENTIST REPORTED THAT AFTER 3 MONTHS DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 7#, IT HAD TO BE REMOVED BECAUSE THE DENTIST WAS NOT ABLE TO REMOVE ONLY THE ABUTMENT, WHICH WAS STUCK INTO THE IMPLANT. IN ADDITION, 50NCM OF PRIMARY STABILITY WAS ACHIEVED, IMMEDIATE IMPLANT WAS PERFORMED AND THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Tissue Damage

    MAUDE report 7483147 ↗
  • InjuryMay 3, 2018Difficult to Remove; Difficult to Remove

    RP 011637- THE DENTIST REPORTED THAT AFTER 3 MONTHS DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 8#, IT HAD TO BE REMOVED BECAUSE THE DENTIST WAS NOT ABLE TO REMOVE ONLY THE ABUTMENT, WHICH WAS STUCK INTO THE IMPLANT. IN ADDITION, 50NCM OF PRIMARY STABILITY WAS ACHIEVED, IMMEDIATE IMPLANT WAS PERFORMED AND THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Tissue Damage

    MAUDE report 7483142 ↗
  • InjuryNov 14, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT, 2 MONTHS AND 2 WEEKS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #5, ITS NON- OSSEOINTEGRATION WAS OBSERVED. FIFTEEN NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED BONE TYPE IV AND BONE GRAFT WAS EXECUTED.

    Patient problems: Tissue Damage; Unspecified Infection

    MAUDE report 7030724 ↗
  • InjuryNov 14, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT, ALMOST 2 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #10, ITS NON-OSSEOINTEGRATION WAS OBSERVED. FORTY FIVE NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/ QUANTITY (BONE TYPE II), FENESTRATION OCCURRED DURING SURGERY AND BONE GRAFT WAS EXECUTED.

    Patient problems: Tissue Damage; Unspecified Infection

    MAUDE report 7029888 ↗
  • InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT, 3 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #7, ITS NON- OSSEOINTEGRATION WAS OBSERVED. THIRTY-FIVE (35) NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Tissue Damage

    MAUDE report 7019447 ↗
  • InjuryNov 10, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    (B)(4) - THE DENTIST REPORTED HAVING TO REMOVE THE DENTAL IMPLANT AFTER ITS OSSEOINTEGRATION, BECAUSE THE ABUTMENT WAS SUBSTITUTED AND THE NEW ABUTMENT GOT STUCK INTO THE IMPLANT. AS IT WAS NOT POSSIBLE TO REMOVE ONLY THE ABUTMENT, THE IMPLANT WAS ALSO REMOVED.

    Patient problems: Tissue Damage

    MAUDE report 7018904 ↗
  • InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 3 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #29, IT WAS VERIFIED ITS NON- OSSEOINTEGRATION. THE PATIENT PRESENTED BONE TYPE III, 45 NCM OF PRIMARY STABILITY WAS ACHIEVED AND IMMEDIATE LOAD WAS NOT PERFORMED.

    Patient problems: Tissue Damage

    MAUDE report 7017089 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →