Skip to content

Neodent 140.756Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · Ti DRIVE ACQUA IMPLANT · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07899878017110), received between Nov 9, 2017 and Jun 29, 2023 — classified as 3 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure to Osseointegrate2
  • Loss of Osseointegration1

Patient problems

  • Tissue Damage2
  • Pain1
  • Swelling/ Edema1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 29, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN FDI 11. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 17231796 ↗
  • InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT, 6 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN PATIENT¿S MOUTH, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THE 30 NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS IMMEDIATELY LOADED. THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Tissue Damage

    MAUDE report 7020698 ↗
  • InjuryNov 9, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 1.5 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #7 IT WAS VERIFIED ITS NON- OSSEOINTEGRATION. 30 N.CM. OF PRIMARY STABILITY WAS ACHIEVED AND IMMEDIATE LOAD WAS NOT PERFORMED. THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Tissue Damage

    MAUDE report 7015902 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page