Skip to content

Neodent 140.757Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · Ti DRIVE ACQUA IMPLANT · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
19
Deaths reported
0
Injuries reported
19
Recalls naming it
0

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 07899878017127), received between Nov 8, 2017 and Feb 7, 2022 — classified as 19 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure to Osseointegrate13
  • Loss of Osseointegration5
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Tissue Damage14
  • Pain5
  • Unspecified Infection3
  • No Clinical Signs, Symptoms or Conditions2
  • Sinus Perforation2
  • Fibrosis1
  • Fistula1

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 7, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 26. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2021 , LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13474738 ↗
  • InjurySep 6, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN FDI 27. DETAILS OF SURGERY: SINUS PERFORATION AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2020, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Sinus Perforation

    MAUDE report 12430058 ↗
  • InjuryMar 15, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2020, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11488017 ↗
  • InjuryNov 30, 2020Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN FDI 25. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2018, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND POOR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION AND PERI-IMPLANTITIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 10922580 ↗
  • InjuryNov 16, 2020Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN FDI 26. DETAILS OF SURGERY: SINUS AUGMENTATION AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2020, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 10846582 ↗
  • InjuryJun 18, 2020Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN FDI 26. DETAILS OF SURGERY: SINUS AUGMENTATION. ON (B)(6) 2020, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: Tissue Damage

    MAUDE report 10169330 ↗
  • InjuryJun 27, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 1 MONTH AND 4 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 14#, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 20N.CM OF PRIMARY STABILITY WAS ACHIEVED, THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY (BONE TYPE IV) AND BONE GRAFT WAS PLACED.

    Patient problems: Tissue Damage

    MAUDE report 8741403 ↗
  • InjuryJun 19, 2019Loss of Osseointegration; Loss of Osseointegration

    (B)(4). THE DENTIST REPORTED THAT ALMOST 1 YEAR AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 19#, ITS LOSS OF OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 45N.CM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY (BONE TYPE II).

    Patient problems: Pain; Tissue Damage

    MAUDE report 8714347 ↗
  • InjuryMar 11, 2019Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) ¿ THE DENTIST REPORTED THAT 1 MONTH AND 7 DAYS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 14#, ITS NON- OSSEOINTEGRATION WAS OBSERVED. MOREOVER, 20N.CM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Fistula; Tissue Damage

    MAUDE report 8407486 ↗
  • InjuryOct 28, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 3 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 14#, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, THE PATIENT PRESENTED BONE TYPE II AND IMMEDIATE IMPLANT PROCEDURE WAS PERFORMED.

    Patient problems: Tissue Damage

    MAUDE report 8011262 ↗
  • InjuryOct 28, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 1 MONTH AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 31#, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, THE PATIENT PRESENTED BONE TYPE II, BONE GRAFT WAS EXECUTED AND IMMEDIATE IMPLANT PROCEDURE WAS PERFORMED.

    Patient problems: Pain; Tissue Damage; Unspecified Infection

    MAUDE report 8011248 ↗
  • InjuryMar 6, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT 1 MONTH AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #23, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THE 45 NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED BONE TYPE I. THE DENTIST ALSO INFORMED THAT IT WAS USED THE PROCEDURE OF IMMEDIATE LOAD.

    Patient problems: Pain; Tissue Damage

    MAUDE report 7317190 ↗
  • InjuryJan 23, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 3 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #6, ITS NON-OSSEOINTEGRATION WAS OBSERVED. THIRTY-THREE (33) NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED BONE TYPE II.

    Patient problems: Tissue Damage

    MAUDE report 7213735 ↗
  • InjuryJan 23, 2018Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 1 MONTH AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #21 ITS NON-OSSEOINTEGRATION WAS OBSERVED. THIRTY (30) NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE IMPLANT WAS INSTALLED WITHOUT IMMEDIATELY LOAD. THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Tissue Damage

    MAUDE report 7213734 ↗
  • InjuryNov 14, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT, 1 MONTH AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #14; ITS NON-OSSEOINTEGRATION WAS OBSERVED. THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY AND BONE GRAFT WAS EXECUTED.

    Patient problems: Sinus Perforation; Tissue Damage; Unspecified Infection

    MAUDE report 7030864 ↗
  • InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT 2 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #14, IT WAS VERIFIED ITS NON- OSSEOINTEGRATION. FORTY NCM OF PRIMARY STABILITY WAS ACHIEVED AND IMMEDIATE LOAD WAS PERFORMED. THE PATIENT PRESENTED BONE TYPE III.

    Patient problems: Tissue Damage

    MAUDE report 7017365 ↗
  • InjuryNov 10, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4) - THE DENTIST REPORTED THAT ALMOST 5 MONTHS AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #14, IT WAS VERIFIED ITS NON- OSSEOINTEGRATION. TEN NCM OF PRIMARY STABILITY WAS ACHIEVED AND IMMEDIATE LOAD WAS NOT PERFORMED. THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/ QUANTITY (BONE TYPE III) AND BONE GRAFT WAS EXECUTED.

    Patient problems: Tissue Damage

    MAUDE report 7016985 ↗
  • InjuryNov 9, 2017Failure to Osseointegrate; Failure to Osseointegrate

    (B)(4). THE DENTIST REPORTED THAT 1 YEAR AFTER THE DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION #14, IT WAS VERIFIED ITS NONOSSEOINTEGRATION. A 10 NCM OF PRIMARY STABILITY WAS ACHIEVED AND IMMEDIATE LOAD WAS NOT PERFORMED. THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY (BONE TYPE IV).

    Patient problems: Tissue Damage

    MAUDE report 7012850 ↗
  • InjuryNov 8, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    OS 111316 - DENTIST REMOVED A DENTAL IMPLANT INSTALLED IN INTRAORAL REGION #29 DUE TO PATIENT'S REQUEST. PATIENT WAS IN PAIN, WHICH HAS CONTINUED EVEN WITH THE IMPLANT REMOVAL.

    Patient problems: Pain; Tissue Damage

    MAUDE report 7010852 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page