Neodent 109.991Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · CM DRIVE IMPLANT · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent CM Drive implant is a root-form implant placed into bone. It has 198 adverse event reports, with no deaths and 198 injuries. Most describe the implant failing to bond with the bone, and some mention gum disease around the implant, bleeding and tissue damage. No reports came in the last 12 months, compared with 6 the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
198
Deaths reported
0
Injuries reported
198
Recalls naming it
0

Summary

The FDA's MAUDE database contains 198 adverse event reports that list this exact device (DI 07899878023302), received between Nov 14, 2017 and Jan 27, 2025 — classified as 198 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (147 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 6 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate147
  • Loss of Osseointegration40
  • Adverse Event Without Identified Device or Use Problem16
  • Fracture2
  • Difficult to Remove1
  • Insufficient Device Problem Information1

Patient problems

  • Tissue Damage129
  • Pain45
  • Unspecified Infection41
  • Hemorrhage/Blood Loss/Bleeding38
  • No Clinical Signs, Symptoms or Conditions17
  • Fibrosis15
  • Inflammation10
  • Failure of Implant9

Recent adverse event reports

The 25 most recent of 198 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 27, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN FDI 14. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21245319 ↗
  • InjuryJan 27, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN FDI 14. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21243407 ↗
  • InjuryJan 3, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN ADA 9. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 21071471 ↗
  • InjuryJan 3, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN ADA 10. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Unspecified Infection

    MAUDE report 21071469 ↗
  • InjuryJan 3, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN ADA 10. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Unspecified Infection

    MAUDE report 21070092 ↗
  • InjuryJan 3, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2017 IN ADA 9. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 21070089 ↗
  • InjuryApr 8, 2024Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2020 IN ADA 23. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19065797 ↗
  • InjuryApr 8, 2024Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN ADA 23. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19062269 ↗
  • InjuryNov 20, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED 2018-(B)(6) IN ADA 19. ON 2023-(B)(6), LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18178131 ↗
  • InjuryNov 20, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018, IN ADA 19. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18173650 ↗
  • InjuryFeb 3, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN ADA 11. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16296460 ↗
  • InjuryFeb 3, 2023Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2019 IN ADA 11. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16295407 ↗
  • InjuryOct 3, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN ADA 3. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND SINUS AUGMENTATION. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15532228 ↗
  • InjuryOct 3, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN ADA 3. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND SINUS AUGMENTATION. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15531198 ↗
  • InjurySep 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 21, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15489633 ↗
  • InjurySep 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 21, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15487606 ↗
  • InjuryMay 20, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 11. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 14464280 ↗
  • InjuryMay 20, 2022Fracture; Fracture

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 10. ON (B)(6) 2022, FRACTURE OF THE IMPLANT WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Inflammation

    MAUDE report 14464206 ↗
  • InjuryMay 20, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 11. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON 2022-02-15, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 14464075 ↗
  • InjuryMay 20, 2022Fracture; Fracture

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 10. ON (B)(6) 2022 FRACTURE OF THE IMPLANT WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Inflammation

    MAUDE report 14463354 ↗
  • InjuryApr 7, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 3. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH POOR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 14046015 ↗
  • InjuryApr 7, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 3. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH POOR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 14045252 ↗
  • InjuryMar 21, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2018 IN THE PATIENT'S MOUTH. ON (B)(6) 2018, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13843534 ↗
  • InjuryNov 11, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 5. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND POOR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 12798407 ↗
  • InjuryNov 11, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 20. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND POOR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 12798356 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →