This Neodent GM Titamax dental implant is a screw-shaped root-form implant used to replace a missing tooth. The 43 reports mostly describe the implant failing to bond with the surrounding bone, so it later becomes loose or falls out; reports also mention infection around the implant, pain, and one broken implant. There are 43 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 44
- Deaths reported
- 0
- Injuries reported
- 44
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 07899878023890), received between Nov 12, 2019 and Sep 7, 2026 — classified as 44 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (40 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate40
- Loss of Osseointegration2
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
Patient problems
- Unspecified Infection16
- Fibrosis15
- No Clinical Signs, Symptoms or Conditions15
- Pain6
- Swelling/ Edema2
- Tissue Damage2
- Numbness1
Recent adverse event reports
The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.
InjurySep 7, 2026Failure to Osseointegrate
Patient problems: Pain; Swelling/ Edema; Unspecified Infection
MAUDE report 50655918 ↗InjuryAug 27, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6)2026 in the patient's mouth. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50482147 ↗InjuryMay 6, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 36. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25102755 ↗InjuryApr 27, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 26. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND FAIR ORAL HYGIENE. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Unspecified Infection
MAUDE report 25007793 ↗InjuryApr 20, 2026Fracture
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 44. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. ON (B)(6) 2026, FRACTURE OF THE IMPLANT WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24931044 ↗InjuryApr 8, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 42. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE AND INADEQUATE BONE QUALITY/QUANTITY. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24819743 ↗InjuryFeb 24, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 25. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025 , NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND FAIR ORAL HYGIENE. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 24432908 ↗InjuryFeb 24, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 45. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND FAIR ORAL HYGIENE. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Unspecified Infection
MAUDE report 24432840 ↗InjuryFeb 3, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 36. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24249697 ↗InjuryFeb 3, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 35. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Pain
MAUDE report 24249692 ↗InjuryJan 6, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 33. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Pain; Unspecified Infection
MAUDE report 23978794 ↗InjuryJan 6, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 14. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Unspecified Infection
MAUDE report 23978777 ↗InjuryDec 30, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 14. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Unspecified Infection
MAUDE report 23926121 ↗InjuryDec 1, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025 , NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23686414 ↗InjuryNov 10, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23517854 ↗InjuryOct 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN FDI 15, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. PATIENT PRESENTED WITH INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 23227110 ↗InjurySep 9, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025 IN FDI 32. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BRUXISM. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: NUMBNESS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Numbness
MAUDE report 23007845 ↗InjuryJul 15, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 17. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND BONE OVERHEATING. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22510773 ↗InjuryJun 30, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22363804 ↗InjuryJun 18, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 35. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain
MAUDE report 22276859 ↗InjuryJun 18, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN FDI 46. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 22263764 ↗InjuryJun 18, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN FDI 46. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22263422 ↗InjuryJun 17, 2025Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN FDI 14. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Unspecified Infection
MAUDE report 22256671 ↗InjuryJun 17, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 17. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Unspecified Infection
MAUDE report 22247516 ↗InjuryJun 17, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 17. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Unspecified Infection
MAUDE report 22247446 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page