Neodent 109.960Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · GM DRIVE IMPLANT · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent GM Drive dental implant has 347 adverse event reports on file, with no deaths and 347 injuries. Most reports describe the implant never fusing with the jawbone or losing its hold, and many mention scar-like tissue, pain or infection around the implant. No recalls name this device. Reports in the last 12 months rose to 86 from 51 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
351
Deaths reported
0
Injuries reported
351
Recalls naming it
0

Summary

The FDA's MAUDE database contains 351 adverse event reports that list this exact device (DI 07899878024316), received between Apr 20, 2018 and Sep 28, 2026 — classified as 351 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (270 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 6 reportsDec 2023: 2 reportsJan 2024: 4 reportsFeb 2024: 8 reportsMar 2024: 3 reportsApr 2024: 6 reportsMay 2024: 8 reportsJun 2024: 8 reportsJul 2024: 11 reportsAug 2024: 1 reportsSep 2024: 6 reportsOct 2024: 4 reportsNov 2024: 5 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 3 reportsMar 2025: 5 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 7 reportsJul 2025: 4 reportsAug 2025: 2 reportsSep 2025: 13 reportsOct 2025: 7 reportsNov 2025: 7 reportsDec 2025: 8 reportsJan 2026: 4 reportsFeb 2026: 13 reportsMar 2026: 13 reportsApr 2026: 10 reportsMay 2026: 11 reportsJun 2026: 11 reportsJul 2026: 4 reportsAug 2026: 4 reportsSep 2026: 5 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate270
  • Loss of Osseointegration46
  • Adverse Event Without Identified Device or Use Problem26
  • Difficult to Insert9

Patient problems

  • No Clinical Signs, Symptoms or Conditions148
  • Fibrosis137
  • Pain88
  • Unspecified Infection77
  • Inflammation29
  • Swelling/ Edema22
  • Hemorrhage/Blood Loss/Bleeding18
  • Tissue Damage17

Recent adverse event reports

The 25 most recent of 351 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50922960 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026 (b)(6) in ADA 14. Details of surgery: Primary stability not achieved. On 2026 (b)(6), non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827468 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50823933 ↗
  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50808108 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50761891 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50417577 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50413407 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted on (b)(6)2026 in FDI 44. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Infection and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Unspecified Infection

    MAUDE report 50284436 ↗
  • InjuryAug 10, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 44. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Peri-implantitis and Pain. No further patient complications were reported.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 50200167 ↗
  • InjuryJul 17, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 12. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25860082 ↗
  • InjuryJul 17, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 26. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Inflammation; Pain; Unspecified Infection

    MAUDE report 25858706 ↗
  • InjuryJul 16, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 24. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND FAIR ORAL HYGIENE. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Unspecified Infection

    MAUDE report 25845409 ↗
  • InjuryJul 10, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 15. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Unspecified Infection

    MAUDE report 25784794 ↗
  • InjuryJun 26, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 6. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV, FAIR ORAL HYGIENE, INADEQUATE BONE QUALITY/QUANTITY AND PREVIOUS BONE AUGMENTATION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 25640249 ↗
  • InjuryJun 26, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 6. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV, FAIR ORAL HYGIENE, INADEQUATE BONE QUALITY/QUANTITY AND PREVIOUS BONE AUGMENTATION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain

    MAUDE report 25639202 ↗
  • InjuryJun 26, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 5. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED, IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY AND FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Fistula

    MAUDE report 25639150 ↗
  • InjuryJun 15, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 26. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 25510013 ↗
  • InjuryJun 15, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 41. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 25509618 ↗
  • InjuryJun 15, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 23. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 25506644 ↗
  • InjuryJun 15, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 26. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN, MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 25506634 ↗
  • InjuryJun 12, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 12. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Unspecified Infection

    MAUDE report 25490482 ↗
  • InjuryJun 12, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 34. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Unspecified Infection

    MAUDE report 25490392 ↗
  • InjuryJun 12, 2026Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 16. ON (B)(6) 2026, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PERI-IMPLANTITIS AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 25490350 ↗
  • InjuryJun 3, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN FDI 35. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25381940 ↗
  • InjuryMay 28, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN ADA 3. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE, BRUXISM AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN, MOBILITY, INCREASED SENSITIVITY, HYPERSENSITIVITY AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hyperresponsive to stimuli; Hypersensitivity/Allergic reaction; Inflammation; Pain

    MAUDE report 25321410 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates