Neodent 109.977Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · GM HELIX IMPLANT · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent GM Helix implant is placed in bone. Reports mostly describe the implant failing to bond with the surrounding bone, along with scar tissue, pain, and infection. There are 6,915 reports, with no deaths and 6,915 injuries. No recalls name this device. Reports rose to 2,349 in the last 12 months from 1,717 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
7,076
Deaths reported
0
Injuries reported
7,076
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7,076 adverse event reports that list this exact device (DI 07899878024484), received between Nov 26, 2018 and Sep 30, 2026 — classified as 7,076 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (6,027 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 104 reportsJan 2024: 54 reportsFeb 2024: 57 reportsMar 2024: 70 reportsApr 2024: 110 reportsMay 2024: 111 reportsJun 2024: 106 reportsJul 2024: 130 reportsAug 2024: 157 reportsSep 2024: 141 reportsOct 2024: 179 reportsNov 2024: 144 reportsDec 2024: 107 reportsJan 2025: 115 reportsFeb 2025: 86 reportsMar 2025: 157 reportsApr 2025: 48 reportsMay 2025: 0 reportsJun 2025: 448 reportsJul 2025: 163 reportsAug 2025: 156 reportsSep 2025: 132 reportsOct 2025: 161 reportsNov 2025: 190 reportsDec 2025: 400 reportsJan 2026: 270 reportsFeb 2026: 191 reportsMar 2026: 249 reportsApr 2026: 207 reportsMay 2026: 227 reportsJun 2026: 184 reportsJul 2026: 171 reportsAug 2026: 240 reportsSep 2026: 181 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026448
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate6,027
  • Loss of Osseointegration689
  • Adverse Event Without Identified Device or Use Problem288
  • Difficult to Insert37
  • Fracture34
  • Separation Failure10

Patient problems

  • No Clinical Signs, Symptoms or Conditions4,290
  • Fibrosis1,435
  • Pain1,362
  • Unspecified Infection1,165
  • Inflammation316
  • Swelling/ Edema289
  • Fistula243
  • Hemorrhage/Blood Loss/Bleeding188

Recent adverse event reports

The 25 most recent of 7,076 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in the patient's mouth. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950383 ↗
  • InjurySep 30, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in the patient's mouth. On (B)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950347 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in ADA 13. On (B)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Pain and Swelling. No further patient complications were reported.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 50923016 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 23. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Pain and Swelling. No further patient complications were reported.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 50923015 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 46. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50922978 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2026(b)(6) in the patient's mouth. On 2026(b)(6), loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50922977 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 22. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922959 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 36. On (B)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50922950 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in FDI 32. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922942 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in the patient's mouth. On (B)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922941 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in ADA 9. On (B)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922897 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2025 in FDI 33. On (B)(6) 2026, non-osseointegration was verified. Patient presented with inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50922843 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922788 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922773 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On 2026-08-25, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922768 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922762 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in the patient's mouth. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922735 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 9. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Mobility, Bleeding, Fistula and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Inflammation; Unspecified Infection

    MAUDE report 50922725 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted on (b)(6) 2026 in ADA 7. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Mobility, Bleeding, Fistula and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Inflammation; Unspecified Infection

    MAUDE report 50922723 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922695 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2025 in the patient's mouth. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III, poor oral hygiene and inadequate bone quality/quantity. No product was returned to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50922650 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    The clinician reports that the day the implant was placed in the patient's mouth, failure occurred upon insertion. Details of surgery: Primary stability not achieved. No product was returned to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922647 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 32. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922618 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922616 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 32. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922615 ↗

FDA clearance and classification

The FDA regulates product code DZE (implant, endosseous, root-form) under 21 CFR 872.3640 as a Class II device. 21 CFR 872.3640 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Neodent 109.977

Has the Neodent 109.977 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07899878024484), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Neodent 109.977?

The FDA's MAUDE database lists 7,076 adverse event reports for this device, received between Nov 26, 2018 and Sep 30, 2026. In those reports, the most common specific device problems are “Failure to Osseointegrate” (6,027), “Loss of Osseointegration” (689) and “Difficult to Insert” (37); the most common patient problems are “Fibrosis” (1,435), “Pain” (1,362) and “Unspecified Infection” (1,165). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Neodent 109.977?

None of the reports are classified as deaths, 7,076 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Neodent 109.977 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Neodent 109.977, and is it still sold?

The FDA's device database lists JJGC Indústria e Comércio de Materiais Dentários S/A. as the company responsible for it (brand Neodent, model 109.977). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.