Neodent 140.946Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · GM HELIX ACQUA IMPLANT · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent GM Helix Acqua dental implant is placed in the jawbone. Reports most often describe the implant failing to fuse with the bone, and some mention scar-like tissue, pain, infection and swelling. The device appears in 5,004 reports, with 5,004 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
5,054
Deaths reported
0
Injuries reported
5,054
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5,054 adverse event reports that list this exact device (DI 07899878024910), received between Oct 4, 2018 and Sep 28, 2026 — classified as 5,054 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (3,300 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 70 reportsJan 2024: 74 reportsFeb 2024: 62 reportsMar 2024: 32 reportsApr 2024: 83 reportsMay 2024: 112 reportsJun 2024: 121 reportsJul 2024: 92 reportsAug 2024: 55 reportsSep 2024: 99 reportsOct 2024: 68 reportsNov 2024: 103 reportsDec 2024: 79 reportsJan 2025: 99 reportsFeb 2025: 64 reportsMar 2025: 99 reportsApr 2025: 25 reportsMay 2025: 0 reportsJun 2025: 200 reportsJul 2025: 103 reportsAug 2025: 138 reportsSep 2025: 55 reportsOct 2025: 73 reportsNov 2025: 118 reportsDec 2025: 115 reportsJan 2026: 84 reportsFeb 2026: 69 reportsMar 2026: 124 reportsApr 2026: 108 reportsMay 2026: 97 reportsJun 2026: 122 reportsJul 2026: 87 reportsAug 2026: 106 reportsSep 2026: 86 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026200
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate3,300
  • Loss of Osseointegration1,469
  • Adverse Event Without Identified Device or Use Problem213
  • Fracture56
  • Difficult to Insert36
  • Separation Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,462
  • Fibrosis1,384
  • Pain1,234
  • Unspecified Infection790
  • Inflammation389
  • Swelling/ Edema258
  • Hemorrhage/Blood Loss/Bleeding237
  • Tissue Damage235

Recent adverse event reports

The 25 most recent of 5,054 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2025(b)(6) in ADA 13. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922939 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2021 in ADA 29. On 2026-08-20, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922924 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2026(b)(6) in ADA 22. Details of surgery: Implant surface not completely covered with bone. On 2026(b)(6), loss of osseointegration was verified. Patient presented with bone type IV, fair oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922921 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 25. Details of surgery: Sinus augmentation and Immediate extraction site. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain, Mobility, Swelling and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Inflammation; Pain; Swelling/ Edema

    MAUDE report 50922906 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 13. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type IV and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50922852 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted on (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Peri-implantitis, Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain; Unspecified Infection

    MAUDE report 50922666 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in ADA 21. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922240 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 12. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922104 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in ADA 13. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918870 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 22. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type IV, fair oral hygiene and inadequate bone quality/quantity. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918851 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2021 in ADA 29. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918815 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 13. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type IV and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50918811 ↗
  • InjurySep 25, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50902255 ↗
  • InjurySep 25, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50902190 ↗
  • InjurySep 25, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 13. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bruxism. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50902120 ↗
  • InjurySep 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 27. On (b)(6) 2025, non-osseointegration was verified. No product was returned to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50888327 ↗
  • InjurySep 24, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 11. Details of surgery: Implant surface not completely covered with bone and Immediate extraction site. On (b)(6) 2026, loss of osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50888317 ↗
  • InjurySep 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 4. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50888226 ↗
  • InjurySep 24, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2026(b)(6) in FDI 32. On 2026(b)(6), loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50879222 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 26. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827864 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 28. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827863 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827733 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 10. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50827548 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 7. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50827537 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 10. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50824196 ↗

FDA clearance and classification

The FDA regulates product code DZE (implant, endosseous, root-form) under 21 CFR 872.3640 as a Class II device. 21 CFR 872.3640 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Neodent 140.946

Has the Neodent 140.946 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07899878024910), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Neodent 140.946?

The FDA's MAUDE database lists 5,054 adverse event reports for this device, received between Oct 4, 2018 and Sep 28, 2026. In those reports, the most common specific device problems are “Failure to Osseointegrate” (3,300), “Loss of Osseointegration” (1,469) and “Fracture” (56); the most common patient problems are “Fibrosis” (1,384), “Pain” (1,234) and “Unspecified Infection” (790). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Neodent 140.946?

None of the reports are classified as deaths, 5,054 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Neodent 140.946 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Neodent 140.946, and is it still sold?

The FDA's device database lists JJGC Indústria e Comércio de Materiais Dentários S/A. as the company responsible for it (brand Neodent, model 140.946). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.