Neodent 140.949Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · GM HELIX ACQUA IMPLANT · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent GM Helix Acqua dental implant is placed in the jawbone to support a replacement tooth. Most reports describe the implant failing to fuse with the bone, sometimes followed by pain, scar tissue or infection around it. There are 3,844 reports, including 3,844 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
3,928
Deaths reported
0
Injuries reported
3,928
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3,928 adverse event reports that list this exact device (DI 07899878024941), received between Feb 7, 2019 and Sep 28, 2026 — classified as 3,928 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (3,012 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 37 reportsJan 2024: 47 reportsFeb 2024: 46 reportsMar 2024: 44 reportsApr 2024: 85 reportsMay 2024: 45 reportsJun 2024: 62 reportsJul 2024: 90 reportsAug 2024: 63 reportsSep 2024: 71 reportsOct 2024: 73 reportsNov 2024: 73 reportsDec 2024: 88 reportsJan 2025: 62 reportsFeb 2025: 68 reportsMar 2025: 55 reportsApr 2025: 40 reportsMay 2025: 0 reportsJun 2025: 222 reportsJul 2025: 97 reportsAug 2025: 95 reportsSep 2025: 75 reportsOct 2025: 99 reportsNov 2025: 83 reportsDec 2025: 135 reportsJan 2026: 97 reportsFeb 2026: 70 reportsMar 2026: 143 reportsApr 2026: 80 reportsMay 2026: 86 reportsJun 2026: 103 reportsJul 2026: 94 reportsAug 2026: 111 reportsSep 2026: 101 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026222
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate3,012
  • Loss of Osseointegration625
  • Adverse Event Without Identified Device or Use Problem239
  • Difficult to Insert54
  • Fracture11
  • Separation Failure3

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,959
  • Pain1,032
  • Fibrosis954
  • Unspecified Infection748
  • Inflammation254
  • Hemorrhage/Blood Loss/Bleeding216
  • Swelling/ Edema194
  • Tissue Damage125

Recent adverse event reports

The 25 most recent of 3,928 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 21. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Pain and Swelling. No further patient complications were reported.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 50923012 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 14. On (b)(6) 2026, loss of osseointegration was verified. No product was returned to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922987 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 4. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility, Swelling, Bleeding, Abscess and Fistula. No further patient complications were reported.

    Patient problems: Abscess; Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Pain; Swelling/ Edema

    MAUDE report 50922938 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    The clinician reports that the day the implant was placed in ADA 19, failure occurred upon insertion. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922877 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922741 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922736 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in the patient's mouth. On 2026-(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922697 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Pain, Mobility and Fistula. No further patient complications were reported.

    Patient problems: Fibrosis; Fistula; Pain

    MAUDE report 50922675 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene. No product was returned to the manufacturer. At the event the patient experienced: Infection, Pain, Mobility and Bleeding. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Pain; Unspecified Infection

    MAUDE report 50922672 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50922668 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50922667 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. No product was returned to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50922663 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50922661 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922346 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922265 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922241 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 4. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility, Swelling, Bleeding, Abscess and Fistula. No further patient complications were reported.

    Patient problems: Abscess; Fibrosis; Fistula; Hemorrhage/Blood Loss/Bleeding; Pain; Swelling/ Edema

    MAUDE report 50918865 ↗
  • InjurySep 25, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50902286 ↗
  • InjurySep 25, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with inadequate bone quality/quantity. No product was returned to the manufacturer. At the event the patient experienced: Infection. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50902154 ↗
  • InjurySep 24, 2026Fracture

    The clinician reports the implant was inserted (b)(6) 2025 in ADA 20. Details of surgery: Implant surface not completely covered with bone, Sinus augmentation and Ridge augmentation. On (b)(6) 2026, fracture of the implant was verified. Patient presented with poor oral hygiene and bruxism. No product was returned to the manufacturer. At the event the patient experienced: Infection, Peri-implantitis, Pain and Inflammation. No further patient complications were reported.

    Patient problems: Inflammation; Pain; Unspecified Infection

    MAUDE report 50888362 ↗
  • InjurySep 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 17. On (b)(6) 2026, non-osseointegration was verified. Patient presented with inadequate bone quality/quantity. No product was returned to the manufacturer. At the event the patient experienced: Pain, Mobility and Swelling. No further patient complications were reported.

    Patient problems: Fibrosis; Pain; Swelling/ Edema

    MAUDE report 50888296 ↗
  • InjurySep 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2026(b)(6) in the patient's mouth. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50888233 ↗
  • InjurySep 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 16. On (b)(6) 2026, non-osseointegration was verified. Patient presented with inadequate bone quality/quantity. No product was returned to the manufacturer. At the event the patient experienced: Pain, Mobility and Swelling. No further patient complications were reported.

    Patient problems: Fibrosis; Pain; Swelling/ Edema

    MAUDE report 50888214 ↗
  • InjurySep 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. Patient presented with inadequate bone quality/quantity. No product was returned to the manufacturer. At the event the patient experienced: Infection, Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain; Unspecified Infection

    MAUDE report 50879127 ↗
  • InjurySep 21, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 13. Details of surgery: Implant surface not completely covered with bone. On( b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility and Abscess. No further patient complications were reported.

    Patient problems: Abscess; Fibrosis; Pain

    MAUDE report 50843660 ↗

FDA clearance and classification

The FDA regulates product code DZE (implant, endosseous, root-form) under 21 CFR 872.3640 as a Class II device. 21 CFR 872.3640 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Neodent 140.949

Has the Neodent 140.949 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07899878024941), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Neodent 140.949?

The FDA's MAUDE database lists 3,928 adverse event reports for this device, received between Feb 7, 2019 and Sep 28, 2026. In those reports, the most common specific device problems are “Failure to Osseointegrate” (3,012), “Loss of Osseointegration” (625) and “Difficult to Insert” (54); the most common patient problems are “Pain” (1,032), “Fibrosis” (954) and “Unspecified Infection” (748). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Neodent 140.949?

None of the reports are classified as deaths, 3,928 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Neodent 140.949 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Neodent 140.949, and is it still sold?

The FDA's device database lists JJGC Indústria e Comércio de Materiais Dentários S/A. as the company responsible for it (brand Neodent, model 140.949). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.