Skip to content

Neodent 140.951Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · GM HELIX ACQUA IMPLANT · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent GM Helix Acqua dental implant has 4,127 adverse event reports, all describing injuries and no deaths. The reports mostly describe implants that fail to fuse with the jawbone or later lose their hold, with patients mentioning pain, infection and tissue damage. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
4,166
Deaths reported
0
Injuries reported
4,166
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4,166 adverse event reports that list this exact device (DI 07899878024965), received between Oct 4, 2018 and Sep 28, 2026 — classified as 4,166 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (2,489 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 43 reportsJan 2024: 53 reportsFeb 2024: 49 reportsMar 2024: 86 reportsApr 2024: 78 reportsMay 2024: 55 reportsJun 2024: 91 reportsJul 2024: 113 reportsAug 2024: 54 reportsSep 2024: 84 reportsOct 2024: 70 reportsNov 2024: 75 reportsDec 2024: 65 reportsJan 2025: 71 reportsFeb 2025: 54 reportsMar 2025: 39 reportsApr 2025: 32 reportsMay 2025: 0 reportsJun 2025: 183 reportsJul 2025: 72 reportsAug 2025: 73 reportsSep 2025: 47 reportsOct 2025: 51 reportsNov 2025: 73 reportsDec 2025: 94 reportsJan 2026: 58 reportsFeb 2026: 69 reportsMar 2026: 92 reportsApr 2026: 45 reportsMay 2026: 77 reportsJun 2026: 70 reportsJul 2026: 44 reportsAug 2026: 88 reportsSep 2026: 65 reportsOct 2026: 0 reportsNov 2026: 0 reports202420252026183
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate2,489
  • Loss of Osseointegration1,344
  • Adverse Event Without Identified Device or Use Problem275
  • Difficult to Insert49
  • Fracture22
  • Separation Failure9

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,921
  • Fibrosis1,224
  • Pain1,073
  • Unspecified Infection812
  • Inflammation330
  • Tissue Damage254
  • Hemorrhage/Blood Loss/Bleeding239
  • Swelling/ Edema187

Recent adverse event reports

The 25 most recent of 4,166 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 14. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922568 ↗
  • InjurySep 25, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 26. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50902395 ↗
  • InjurySep 24, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 46. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility, Bleeding and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation

    MAUDE report 50888416 ↗
  • InjurySep 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 26. Details of surgery: Primary stability not achieved, Implant surface not completely covered with bone and Immediate extraction site. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and inadequate bone quality/quantity. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Mobility and Bleeding. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Unspecified Infection

    MAUDE report 50888270 ↗
  • InjurySep 24, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 32. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility, Bleeding and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation

    MAUDE report 50888235 ↗
  • InjurySep 24, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 36. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility, Bleeding and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation

    MAUDE report 50888232 ↗
  • InjurySep 21, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2026(b)(6) in ADA 10. Details of surgery: Primary stability not achieved. On 2026(b)(6), loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Peri-implantitis and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Unspecified Infection

    MAUDE report 50843546 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 45. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827855 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis and Inflammation. No further patient complications were reported.

    Patient problems: Inflammation; Unspecified Infection

    MAUDE report 50827840 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 34. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50827669 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 35. On (b)(6) 2026, non-osseointegration was verified. No product was returned to the manufacturer. At the event the patient experienced: Pain. No further patient complications were reported.

    Patient problems: Pain

    MAUDE report 50827640 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type IV. The device was forwarded to the manufacturer. At the event the patient experienced: Infection, Pain, Swelling and Inflammation. No further patient complications were reported.

    Patient problems: Inflammation; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 50827589 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 4. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827581 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 23. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827579 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 26. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827576 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 13. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827575 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 7. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827569 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 13. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827568 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 4. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50827556 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 20. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis. No further patient complications were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50827546 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2025(b)(6) in FDI 48. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50827136 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 46. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Mobility. No further patient complications were reported.

    Patient problems: Fibrosis

    MAUDE report 50827132 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 4. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50824297 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 23. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50824279 ↗
  • InjurySep 18, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 26. On (b)(6) 2026, loss of osseointegration was verified. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50824272 ↗

FDA clearance and classification

The FDA regulates product code DZE (implant, endosseous, root-form) under 21 CFR 872.3640 as a Class II device. 21 CFR 872.3640 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Neodent 140.951

Has the Neodent 140.951 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07899878024965), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Neodent 140.951?

The FDA's MAUDE database lists 4,166 adverse event reports for this device, received between Oct 4, 2018 and Sep 28, 2026. In those reports, the most common specific device problems are “Failure to Osseointegrate” (2,489), “Loss of Osseointegration” (1,344) and “Difficult to Insert” (49); the most common patient problems are “Fibrosis” (1,224), “Pain” (1,073) and “Unspecified Infection” (812). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Neodent 140.951?

None of the reports are classified as deaths, 4,166 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Neodent 140.951 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Neodent 140.951, and is it still sold?

The FDA's device database lists JJGC Indústria e Comércio de Materiais Dentários S/A. as the company responsible for it (brand Neodent, model 140.951). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page