This Neodent GM Zygomatic Implant is a dental implant used for the upper back jaw bone. Most of the 28 reports describe the implant failing to fuse with the surrounding bone, sometimes after it had already fused, with patients reporting pain, infection, swelling and a loose implant;28 injuries and no deaths were reported. Reports dropped from 11 in the earlier 12 months to 5 in the most recent 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 27
- Deaths reported
- 0
- Injuries reported
- 27
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 07899878032038), received between Jan 3, 2022 and May 6, 2026 — classified as 27 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (18 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate18
- Loss of Osseointegration4
- Adverse Event Without Identified Device or Use Problem3
- Fracture2
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Pain8
- Fibrosis7
- Unspecified Infection7
- Fistula2
- Swelling/ Edema2
- Inflammation1
Recent adverse event reports
The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.
InjuryMay 6, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH INADEQUATE BONE QUALITY/QUANTITY. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Pain; Unspecified Infection
MAUDE report 25103692 ↗InjuryMar 30, 2026Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2026 IN THE PATIENT'S MOUTH. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN, SWELLING AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Inflammation; Pain; Swelling/ Edema; Unspecified Infection
MAUDE report 24735221 ↗InjuryJan 26, 2026Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 23. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 24182393 ↗InjuryNov 13, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6)2025 IN ADA 16. ON (B)(6)2025, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Unspecified Infection
MAUDE report 23546836 ↗InjuryJul 21, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 23. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22559076 ↗InjuryJul 21, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 12. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22559067 ↗InjuryJun 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN FDI 13, FAILURE OCCURRED UPON INSERTION. PATIENT PRESENTED WITH BONE TYPE III. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22317346 ↗InjuryJun 18, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 25. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22269186 ↗InjuryJun 18, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 15. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22269141 ↗InjuryJun 17, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain
MAUDE report 22229051 ↗InjuryJun 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 10, FAILURE OCCURRED UPON INSERTION. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 22228650 ↗InjuryJun 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 10, FAILURE OCCURRED UPON INSERTION. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 22223347 ↗InjuryJun 17, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Unspecified Infection
MAUDE report 22220641 ↗InjuryMar 20, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2024 IN FDI 26. DETAILS OF SURGERY: RIDGE AUGMENTATION. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis; Pain
MAUDE report 21657115 ↗InjuryFeb 19, 2025Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21413637 ↗InjuryOct 28, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN THE PATIENT'S MOUTH. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 20549505 ↗InjuryAug 12, 2024Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2023 IN FDI 15. ON (B)(6) 2024, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19965073 ↗InjuryJan 22, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN THE PATIENT'S MOUTH. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18559774 ↗InjuryJan 2, 2024Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 23. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18431868 ↗InjuryOct 26, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 15. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18015967 ↗InjuryOct 16, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 36. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain; Unspecified Infection
MAUDE report 17946315 ↗InjuryMar 23, 2023Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN THE PATIENT'S MOUTH. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain; Swelling/ Edema
MAUDE report 16605134 ↗InjuryFeb 27, 2023Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2020 IN ADA 13. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND SINUS AUGMENTATION. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN AND FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fistula; Pain; Unspecified Infection
MAUDE report 16447758 ↗InjuryFeb 27, 2023Loss of Osseointegration; Loss of Osseointegration
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2020 IN ADA 13. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND SINUS AUGMENTATION. ON (B)(6) 2023, LOSS OF OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN AND FISTULA. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fistula; Pain; Unspecified Infection
MAUDE report 16445110 ↗InjuryMar 31, 2022Failure to Osseointegrate; Failure to Osseointegrate
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN ADA 6. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Fibrosis
MAUDE report 13968711 ↗
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,616 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,404 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,400 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,621 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,298 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,448 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,320 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,267 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page