Neodent 140.1060Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · GM HELIX ACQUA IMPLANT · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Neodent GM Helix Acqua implant is a root-form dental implant used to replace missing teeth. The 47 reports mostly describe implants that fail to fuse with the surrounding bone, sometimes loosening and becoming mobile or losing bone integration months after placement, along with pain, swelling, infection and bleeding. No deaths were reported; all 47 reports describe injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
47
Deaths reported
0
Injuries reported
47
Recalls naming it
0

Summary

The FDA's MAUDE database contains 47 adverse event reports that list this exact device (DI 07899878044529), received between Mar 8, 2022 and Aug 13, 2026 — classified as 47 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (23 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 5 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 5 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 4 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate23
  • Adverse Event Without Identified Device or Use Problem16
  • Difficult to Insert4
  • Loss of Osseointegration4

Patient problems

  • Fibrosis20
  • No Clinical Signs, Symptoms or Conditions20
  • Pain12
  • Unspecified Infection8
  • Hemorrhage/Blood Loss/Bleeding4
  • Swelling/ Edema3
  • Hypersensitivity/Allergic reaction2
  • Inflammation2

Recent adverse event reports

The 25 most recent of 47 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 13, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in ADA 12. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50240072 ↗
  • InjuryAug 13, 2026Loss of Osseointegration

    The clinician reports the implant was inserted 2026 (b)(6) in ADA 12. Details of surgery: Implant surface not completely covered with bone. On 2026 (b)(6), loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Mobility. No further patient complications were reported.

    Patient problems: Fibrosis; Pain

    MAUDE report 50238100 ↗
  • InjuryMay 26, 2026Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 2, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. PATIENT PRESENTED WITH BONE TYPE IV AND FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25283659 ↗
  • InjuryMay 26, 2026Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 2, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. PATIENT PRESENTED WITH BONE TYPE IV AND FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25281554 ↗
  • InjuryMay 15, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 3. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25190694 ↗
  • InjuryMay 15, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025 IN ADA 3. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis

    MAUDE report 25188099 ↗
  • InjuryApr 27, 2026Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 19, FAILURE OCCURRED UPON INSERTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25007318 ↗
  • InjuryApr 27, 2026Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 19, FAILURE OCCURRED UPON INSERTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25002965 ↗
  • InjuryFeb 24, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 25. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE AND SINUS PERFORATION. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN, MOBILITY AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain; Sinus Perforation; Swelling/ Edema; Unspecified Infection

    MAUDE report 24433056 ↗
  • InjuryJan 19, 2026Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 4, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY, BLEEDING AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation

    MAUDE report 24110973 ↗
  • InjuryJan 19, 2026Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 4, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED. PATIENT PRESENTED WITH BONE TYPE III AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY, BLEEDING AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding; Inflammation

    MAUDE report 24109620 ↗
  • InjuryJan 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24057884 ↗
  • InjuryJan 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24057883 ↗
  • InjuryJan 13, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24057870 ↗
  • InjuryOct 23, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 46. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN AND SWELLING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 23371771 ↗
  • InjuryAug 4, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 15. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 22707441 ↗
  • InjuryAug 4, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 15. DETAILS OF SURGERY: IMMEDIATE EXTRACTION SITE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 22702528 ↗
  • InjuryJul 4, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 19. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain; Unspecified Infection

    MAUDE report 22405187 ↗
  • InjuryJul 4, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN ADA 19. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Pain; Unspecified Infection

    MAUDE report 22404178 ↗
  • InjuryJun 24, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 26. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN AND HYPERSENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hypersensitivity/Allergic reaction; Pain; Unspecified Infection

    MAUDE report 22317381 ↗
  • InjuryJun 18, 2025Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED ON (B)(6) 2025 IN FDI 35. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS VERIFIED. NO PRODUCT WAS RETURNED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PAIN, SWELLING AND HYPERSENSITIVITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Hypersensitivity/Allergic reaction; Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 22281494 ↗
  • InjuryJun 18, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 17. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMMEDIATE EXTRACTION SITE. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE IV AND INADEQUATE BONE QUALITY/QUANTITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain

    MAUDE report 22264456 ↗
  • InjuryJun 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 31, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22248938 ↗
  • InjuryJun 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN ADA 31, FAILURE OCCURRED UPON INSERTION. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY AND BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22248042 ↗
  • InjuryMar 17, 2025Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 24. ON (B)(6) 2024, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION, PERI-IMPLANTITIS AND MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Fibrosis; Unspecified Infection

    MAUDE report 21625282 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices