Skip to content

Neodent 103.619Recalls & reported problems

JJGC Indústria e Comércio de Materiais Dentários S/A. · LATERAL CUT DRILL ZYGOMA-S · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07899878059820), received between Aug 18, 2026 and Aug 18, 2026 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Foreign Body In Patient1
  • Inflammation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 18, 2026Adverse Event Without Identified Device or Use Problem

    A complaint was received n which the customer reported an intraoperative incident involving item 103.619 (Multifluted Zygoma-S Bur, lot YKWX5). According to the practitioner's report, the multifluted bur fractured, resulting in the bur becoming lodged within the patient's left maxillary sinus. However, the Product Safety Management and Product Research & Development departments conducted the relevant investigations and concluded that the item's drilling guide pin had unscrewed and detached.

    Patient problems: Foreign Body In Patient; Inflammation

    MAUDE report 50307340 ↗

Other DRILL, DENTAL, INTRAORAL devices

All DRILL, DENTAL, INTRAORAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page