Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08022057012739), received between Nov 14, 2019 and Jul 26, 2022 — classified as 8 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information6
- Adverse Event Without Identified Device or Use Problem4
- Difficult to Open or Close2
Patient problems
- Insufficient Information4
- Unspecified Infection4
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJul 26, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
THE MANUFACTURER WAS INFORMED OF THIS EVENT THROUGH THE PATIENT TRACKING DEPARTMENT. BASED ON THE INFORMATION RECEIVED, A CARBOMEDICS MECHANICAL STANDARD HEART VALVE M7-029 WAS IMPLANTED IN THE PATIENT ON (B)(6) 2022. BASED ON THE INFORMATION ON PATIENT'S IMPLANT FORM, ON (B)(6) 2022, A SIMILAR DEVICE M7-029 WAS IMPLANTED IN THE PATIENT. THE MANUFACTURER HAS NOT RECEIVED ANY ALLEGATION OF DEVICE MALFUNCTION NOR OF SERIOUS INJURY ON THE PATIENT FROM THE SITE REGARDING THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 15106052 ↗InjuryJul 26, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
THE MANUFACTURER WAS INFORMED OF THIS EVENT THROUGH THE PATIENT TRACKING DEPARTMENT. BASED ON THE INFORMATION RECEIVED, A CARBOMEDICS MECHANICAL STANDARD HEART VALVE M7-029 WAS IMPLANTED IN THE PATIENT ON (B)(6) 2022. BASED ON THE INFORMATION ON PATIENT'S IMPLANT FORM, ON (B)(6) 2022, A SIMILAR DEVICE M7-029 WAS IMPLANTED IN THE PATIENT. THE MANUFACTURER HAS NOT RECEIVED ANY ALLEGATION OF DEVICE MALFUNCTION NOR OF SERIOUS INJURY ON THE PATIENT FROM THE SITE REGARDING THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 15106046 ↗InjuryMay 25, 2021Difficult to Open or Close; Difficult to Open or Close
ON (B)(6) 2021, A CARBOMEDICS STANDARD MITRAL VALVE M7-029 WAS IMPLANTED IN A PATIENT. AFTER IMPLANTATION, VASCULAR ULTRASOUND (TEE) FOUND THAT ONE OF THE LEAFLETS WAS NOT FLEXIBLE MOVED ENOUGH. CONSIDERING THAT THE PATIENT HAD NOT RECEIVED THE ORIGINAL LEAFLET TREATMENT, AND ONE LEAFLET WAS HYPERTROPHY, THE PHYSICIAN REOPENED THE LEAFLET AND ROTATED THE VALVE IN ONE DIRECTION (ABOUT 30 DEGREES COUNTERCLOCKWISE), BUT THE VASCULAR ULTRASOUND OBSERVATION SHOWED THAT THE LEAFLET WAS STILL INFLEXIBLE AND AFFECTED BLOOD FLOW. AS CLARIFIED, DURING TESTING WITH THE DEDICATED LEAFLET TESTER, ONE LEAFLET WAS MOVING NOT VERY SMOOTHLY, AND A RESISTANCE FORCE WAS NOTED DURING THE OPEN AND CLOSE. THE VALVE WAS ULTIMATELY REPLACED WITH A NEW VALVE FROM ANOTHER BRAND (MEDTRONIC #29). AS REPORTED, THE PATIENT HAD INFECTIVE ENDOCARDITIS WITH THICKENED LEAFLET (ANTERIOR EXTERNAL JUNCTION), BUT THE RESECTION OF THE NATIVE MITRAL VALVE DID NOT AFFECT THE FUNCTION OF THE IMPLANTED VALVE. THE PATIENT REMAINED STABLE DURING THE PROCEDURE WHICH WAS COMPLETED UNEVENTFULLY AFTER THE NEW VALVE WAS IMPLANTED. THE PATIENT RECOVERS WELL.
Patient problems: Insufficient Information
MAUDE report 11881853 ↗InjuryMay 25, 2021Difficult to Open or Close; Difficult to Open or Close
ON (B)(6) 2021, A CARBOMEDICS STANDARD MITRAL VALVE M7-029 WAS IMPLANTED IN A PATIENT. AFTER IMPLANTATION, VASCULAR ULTRASOUND FOUND THAT ONE OF THE LEAFLETS WAS NOT FLEXIBLE MOVED ENOUGH. CONSIDERING THAT THE PATIENT HAD NOT RECEIVED THE ORIGINAL LEAFLET TREATMENT, AND ONE LEAFLET WAS HYPERTROPHY, THE PHYSICIAN REOPENED THE LEAFLET AND ROTATED THE VALVE IN ONE DIRECTION, BUT THE VASCULAR ULTRASOUND OBSERVATION SHOWED THAT THE LEAFLET WAS STILL INFLEXIBLE AND AFFECTED BLOOD FLOW. THE VALVE WAS ULTIMATELY REPLACED WITH A NEW VALVE FROM ANOTHER BRAND (DETAILS UNKNOWN).
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11881772 ↗InjuryNov 14, 2019Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
BASED ON THE ASSESSMENT RECEIVED FROM THE HOSPITAL STAFF, THERE WAS NO EVIDENCE IDENTIFIED TO ATTRIBUTE THE PATIENT'S INFECTION TO THE LIVANOVA DEVICE, WHICH RELATIONSHIP WITH THE EVENT WAS EXCLUDED. THE REMAINDER OF THE INFORMATION PREVIOUSLY SUBMITTED REMAINS UNCHANGED.
Patient problems: Unspecified Infection
MAUDE report 9325034 ↗InjuryNov 14, 2019Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
BASED ON THE ASSESSMENT RECEIVED FROM THE HOSPITAL STAFF, THERE WAS NO EVIDENCE IDENTIFIED TO ATTRIBUTE THE PATIENT'S INFECTION TO THE LIVANOVA DEVICE, WHICH RELATIONSHIP WITH THE EVENT WAS EXCLUDED. THE REMAINDER OF THE INFORMATION PREVIOUSLY SUBMITTED REMAINS UNCHANGED.
Patient problems: Unspecified Infection
MAUDE report 9324982 ↗InjuryNov 14, 2019Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
ON (B)(6) 2019, A PATIENT RECEIVED A CARBOMEDICS STANDARD MITRAL M7-029 (FOURTH CARDIAC SURGERY FOR THE PATIENT). IN THE ICU, A SERIOUS INFECTION (NOT FURTHER SPECIFIED) WAS OBSERVED IN THE PATIENT. A POTENTIAL LINK WITH THE METAL VALVE (MECHANICAL) WAS SUSPECTED, ALTHOUGH THIS RELATIONSHIP WAS NOT FURTHER SPECIFIED NOR SUPPORTED BY FURTHER DATA. THE PATIENT IS REPORTEDLY BEING MONITORED. PER ADDITIONAL INFORMATION RECEIVED, THE PATIENT IS STILL HOSPITALIZED IN THE ICU IN (B)(6) 2019 (MORE THAN 115 DAYS). THE PATIENT HAS TRACHEAL STENOSIS, AND THE PULMONARY PART IS GENERALLY IMPAIRED.
Patient problems: Unspecified Infection
MAUDE report 9324909 ↗InjuryNov 14, 2019Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
ON (B)(6) 2019, A PATIENT RECEIVED A CARBOMEDICS STANDARD MITRAL M7-029. DURING THE SAME PROCEDURE, THE PATIENT RECEIVED ALSO A CARBOMEDICS STANDARD AORTIC A5-021 (B)(4), SUBMITTED SEPARATELY). THIS IS REPORTED AS THE PATIENT'S FOURTH CARDIAC SURGERY. IN THE ICU, A SERIOUS INFECTION (NOT FURTHER SPECIFIED) WAS OBSERVED IN THE PATIENT. A POTENTIAL LINK WITH THE METAL VALVE (MECHANICAL) WAS INITIALLY SUSPECTED, ALTHOUGH THIS RELATIONSHIP WAS NOT FURTHER SPECIFIED NOR SUPPORTED BY FURTHER DATA. THE PATIENT IS REPORTEDLY BEING MONITORED. PER ADDITIONAL INFORMATION RECEIVED, THE PATIENT IS STILL HOSPITALIZED IN THE ICU ON (B)(6) 2019 (MORE THAN 115 DAYS). THE PATIENT HAS TRACHEAL STENOSIS, AND THE PULMONARY PART IS GENERALLY IMPAIRED. BASED ON THE ASSESSMENT RECEIVED FROM THE HOSPITAL STAFF, THERE WAS NO EVIDENCE IDENTIFIED TO ATTRIBUTE THE PATIENT'S INFECTION TO THE LIVANOVA DEVICE, WHICH RELATIONSHIP WITH THE EVENT WAS EXCLUDED.
Patient problems: Unspecified Infection
MAUDE report 9324829 ↗
Other HEART-VALVE, MECHANICAL devices
- On-X Mitral Heart Valve with Conform-X Sewing Ring ONXMC-25/33On-X Life Technologies, Inc. · Heart-Valve, Mechanical1 recall1 death34 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths70 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL5 deaths66 reports
- SJM Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths44 reports
- CARBOMEDICS SUPRA-ANNULAR (TOP HAT) S5-021Corcym Srl · HEART-VALVE, MECHANICAL7 deaths37 reports
- Regent™ 21AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths35 reports
- Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL2 deaths33 reports
- Regent™ 19AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL31 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page